Measurement method for roll angular displacement with a high resolution by using diffraction gratings and a heterodyne interferometer
Authors:
- 1 April 2014
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 85 (4) , 045110
Abstract
The roll angle measurement is difficult to be achieved directly using a typical commercial interferometer due to its low sensitivity in axial direction, where the axial direction is orthogonal to the plane of the roll angular displacement. A roll angle measurement method combined diffraction gratings with a laser heterodyne interferometer is discussed in this paper. The diffraction grating placed in the plane of a roll angular displacement and the interferometer arranged in the plane's orthogonal direction, constitute the measurement pattern for the roll angle with high resolution. The roll angular displacement, considered as the linear, can be tested precisely when the corresponding angle is very small. Using the proposed method, the angle roll measurement obtains the high resolution of 0.002″. Experiment has proved its feasibility and practicability.Keywords
This publication has 12 references indexed in Scilit:
- Small angle measurement method based on the total internal multi-reflectionOptics & Laser Technology, 2012
- Error analysis of a plane mirror interferometer based on geometric optical pathsOptics Express, 2012
- Linear diffraction grating interferometer with high alignment tolerance and high accuracyApplied Optics, 2011
- A linear diffraction grating interferometer with high accuracy2006
- Application of a diffraction grating and position sensitive detectors to the measurement of error motion and angular indexing of an indexing tablePrecision Engineering, 2005
- Roll angular displacement measurement system with microradian accuracySensors and Actuators A: Physical, 2004
- Total internal reflection for precision small-angle measurementApplied Optics, 2001
- Michelson interferometer for precision angle measurementApplied Optics, 1999
- Technology and applications of grating interferometers in high-precision measurementPrecision Engineering, 1992
- High-precision, wide-range, dual-axis, angle monitoring systemReview of Scientific Instruments, 1983