Inelastic X-ray scattering with 0.75 meV resolution at 25.7 keV using a temperature-gradient analyzer
Open Access
- 1 January 2015
- journal article
- research article
- Published by International Union of Crystallography (IUCr) in Journal of Synchrotron Radiation
- Vol. 22 (1) , 3-9
Abstract
The use of temperature-gradient analyzers for non-resonant high-resolution inelastic X-ray scattering is investigated. The gradient compensates for geometrical broadening of the energy resolution by adjusting the lattice spacing of the analyzer crystal. Applying a ∼12 mK temperature gradient across a 9.5 cm analyzer, resolutions of 0.75 (2) meV FWHM at 25.7 keV for Si(13 13 13) and 1.25 (2) meV at 21.7 keV for Si(11 11 11) were measured, while retaining large (250 mm) clearance between the sample position and detector, and reasonable (9.3 mrad × 8.8 mrad) analyzer acceptance. The temperature control and stability are discussed.Keywords
This publication has 4 references indexed in Scilit:
- New developments in fabrication of high-energy-resolution analyzers for inelastic X-ray spectroscopyJournal of Synchrotron Radiation, 2011
- Temperature gradient analyzers for compact high-resolution X-ray spectrometersJournal of Synchrotron Radiation, 2009
- Precise determination of elastic constants by high-resolution inelastic X-ray scatteringJournal of Synchrotron Radiation, 2008