Wavefront Analyzers Induce Instrument Myopia
Authors:
- Alejandro CervinoAlejandro Cervino
- Sarah L HoskingSHSarah L Hosking
- Gurjeet K RaiGRGurjeet K Rai
- Shezhad A NarooSNShezhad A Naroo
- Bernard GilmartinBGBernard Gilmartin
- 1 October 2006
- journal article
- research article
- Published by SLACK, Inc. in Journal of Refractive Surgery
- Vol. 22 (8) , 795-803
Abstract
PURPOSE: To assess the accuracy of three wavefront analyzers versus a validated binocular open-view autorefractor in determining refractive error in non-cycloplegic eyes. METHODS: Eighty eyes were examined using the SRW-5000 open-view infrared autorefractor and, in randomized sequence, three wavefront analyzers: 1) OPD-Scan (NIDEK, Gamagori, Japan), 2) WASCA (Zeiss/Med itec, Jena, Germany), and 3) Allegretto (WaveLight Laser Technologies AG, Erlangen, Germany). Subjects were healthy adults (19 men and 21 women; mean age: 20.8±2.5 years). Refractive errors ranged from +1.5 to -9.75 diopters (D) (mean: +1.83±2.74 D) with up to 1.75 D cylinder (mean: 0.58±0.53 D). Three readings were collected per instrument by one examiner without anticholinergic agents. Refraction values were decomposed into vector components for analysis, resulting in mean spherical equivalent refraction (M) and J0 and J45 being vectors of cylindrical power at 0° and 45°, respectively. RESULTS: Positive correlation was observed between wavefront analyzers and the SRW-5000 for spherical equivalent refraction (OPD-Scan, r=0.959, P<.001; WASCA, r=0.981, P<.001; Allegretto, r=0.942, P<.001). Mean differences and limits of agreement showed more negative spherical equivalent refraction with wavefront analyzers (OPD-Scan, 0.406+0.768 D [range: 0.235 to 0.580 D] [P<.001]; WASCA, 0.511±0.550 D [range: 0.390 to 0.634 D] [P<.001]; and Allegretto, 0.434±0.904 D [range: 0.233 to 0.635 D] [P<.001]). A second analysis eliminating outliers showed the same trend but lower differences: OPD-Scan (n=75), 0.24±0.41 D (range: 0.15 to 0.34 D) (P<.001); WASCA (n=78), 0.46+0.47 D (range: 0.36 to 0.57 D) (P<.001); and Allegretto (n=77), 0.30+0.62 D (range: 0.16 to 0.44 D) (P<.001). No statistically significant differences were noted for J p 0 and J 45 . CONCLUSIONS: Wavefront analyzer refraction resulted in 0.30 D more myopia compared to SRW-5000 refraction in eyes without cycloplegia. This is the result of the accommodation excess attributable to instrument myopia. For the relatively low degrees of astigmatism in this study (<2.0 D), good agreement was noted between wavefront analyzers and the SRW-5000. [J Refract Surg. 2006;22:795-803.]Keywords
This publication has 36 references indexed in Scilit:
- Incidence and Progression of Myopia in Singaporean School ChildrenInvestigative Ophthalmology & Visual Science, 2005
- Accuracy and precision of objective refraction from wavefront aberrationsJournal of Vision, 2004
- A population study on changes in wave aberrations with accommodationJournal of Vision, 2004
- Test–Retest Reliability of Clinical Shack-Hartmann MeasurementsInvestigative Ophthalmology & Visual Science, 2004
- Validation of a Clinical Shack-Hartmann AberrometerOptometry and Vision Science, 2003
- Measurement of Refractive Errors in Young Myopes Using the COAS Shack-Hartmann AberrometerOptometry and Vision Science, 2003
- Measuring of ocular wavefront aberration in large pupils using OPD-scanSeminars in Ophthalmology, 2003
- Ocular optical aberrometer for clinical useJournal of Biomedical Optics, 2001
- Reversal of tropicamide mydriasis with single instillations of pilocarpine can induce substantial pseudo‐myopia in young adultsOphthalmic and Physiological Optics, 1995
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986