Evaluation of Immunoassays and General Biological Indicator Tests for Field Screening of Bacillus anthracis and Ricin
Authors:
- Rachel A. BartholomewRBRachel A. Bartholomew
- Richard M. OzanichRORichard M. Ozanich
- Jennifer S. ArceJAJennifer S. Arce
- Heather E. EngelmannHEHeather E. Engelmann
- Alejandro Heredia-LangnerAHAlejandro Heredia-Langner
- Beth A. HofstadBHBeth A. Hofstad
- Janine R. HutchisonJHJanine R. Hutchison
- Kristin JarmanKJKristin Jarman
- Angela M. MelvilleAMAngela M. Melville
- Kristin D. VictryKVKristin D. Victry
- Cynthia J. Bruckner-LeaCBCynthia J. Bruckner-Lea
Open Access
- 1 February 2017
- journal article
- research article
- Published by SAGE Publications in Health Security
- Vol. 15 (1) , 81-96
Abstract
There is little published data on the performance of biological indicator tests and immunoassays that could be used by first responders to determine if a suspicious powder contains a potential biothreat agent. We evaluated a range of biological indicator tests, including 3 protein tests, 2 ATP tests, 1 DNA test, and 1 FTIR spectroscopy instrument for their ability to screen suspicious powders for Bacillus anthracis ( B. anthracis ) spores and ricin. We also evaluated 12 immunoassays (mostly lateral flow immunoassays) for their ability to screen for B. anthracis and ricin. We used a cost-effective, statistically based test plan that allows instruments to be evaluated at performance levels ranging from 0.85 to 0.95 lower confidence bound of the probability of detection at confidence levels of 80% to 95%. We also assessed interference with 22 common suspicious powders encountered in the field. The detection reproducibility for the biological indicators was evaluated at 10 8 B. anthracis spores and 62.5 μg ricin, and the immunoassay detection reproducibility was evaluated at 10 7 spores/mL ( B. anthracis ) and 0.1 μg/mL (ricin). Seven out of 12 immunoassays met our most stringent criteria for B. anthracis detection, while 9 out of 12 met our most stringent test criteria for ricin detection. Most of the immunoassays also detected ricin in 3 different crude castor seed preparations. Our testing results varied across products and sample preparations, indicating the importance of reviewing performance data for specific instruments and sample types of interest for the application in order to make informed decisions regarding the selection of biodetection equipment for field use.Keywords
This publication has 24 references indexed in Scilit:
- Comparison of eleven commercially available rapid tests for detection of Bacillus anthracis, Francisella tularensis and Yersinia pestisLetters in Applied Microbiology, 2015
- Evaluating 6 Ricin Field Detection AssaysBiosecurity and Bioterrorism: Biodefense Strategy, Practice, and Science, 2014
- Comprehensive Laboratory Evaluation of a Highly Specific Lateral Flow Assay for the Presumptive Identification of Ricin in Suspicious White Powders and Environmental SamplesBiosecurity and Bioterrorism: Biodefense Strategy, Practice, and Science, 2013
- Amperometric Detection ofBacillus anthracisSpores: A Portable, Low-Cost Approach to the ELISAInternational Journal of Electrochemistry, 2013
- Integration of gas chromatography mass spectrometry methods for differentiating ricin preparation methodsThe Analyst, 2012
- An evaluation of suspicious powder screening tools for first respondersJournal of Hazardous Materials, 2009
- Analysis of broth-culturedBacillus atrophaeusandBacillus cereussporesJournal of Applied Microbiology, 2008
- Evaluation of the rapid analyte measurement platform (RAMP) for the detection of Bacillus anthracis at a crime sceneForensic Science International, 2007
- Evaluation of lateral flow assays for the detection of botulinum neurotoxin type A and their application in laboratory diagnosis of botulismDiagnostic Microbiology and Infectious Disease, 2007
- Performance Assessment of Three Commercial Assays for Direct Detection of Bacillus anthracis SporesJournal of Clinical Microbiology, 2003