32‐Channel RF coil optimized for brain and cervical spinal cord at 3 T
Open Access
- 23 March 2011
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 66 (4) , 1198-1208
Abstract
Diffusion and functional magnetic resonance imaging of the spinal cord remain challenging due to the small cross‐sectional size of the cord and susceptibility‐related distortions. Although partially addressable through parallel imaging, few highly parallel array coils have been implemented for the cervical cord. Here, we developed a 32‐channel coil that fully covers the brain and c‐spine and characterized its performance in comparison with a commercially available head/neck/spine array. Image and temporal signal‐to‐noise ratio were, respectively, increased by 2× and 1.8× in the cervical cord. Averaged g‐factors at 4× acceleration were lowered by 22% in the brain and by 39% in the spinal cord, enabling 1‐mm isotropic R = 4 multi‐echo magnetization prepared gradient echo of the full brain and c‐spine in 3:20 min. Diffusion imaging of the cord at 0.6 × 0.6 × 5 mm3 resolution and tractography of the full brain and c‐spine at 1.7‐mm isotropic resolution were feasible without noticeable distortion. Improvements of this nature potentially enhance numerous basic and clinical research studies focused on spinal and supraspinal regions. Magn Reson Med, 2011.Keywords
Funding Information
- National Institute of Health (P41RR14075, U01MH093765)
- French MS Society (ARSEP)
This publication has 35 references indexed in Scilit:
- BOLD signal responses to controlled hypercapnia in human spinal cordNeuroImage, 2010
- 96‐Channel receive‐only head coil for 3 Tesla: Design optimization and evaluationMagnetic Resonance in Medicine, 2009
- Diffusion Spectrum Imaging Shows the Structural Basis of Functional Cerebellar Circuits in the Human Cerebellum In VivoPLOS ONE, 2009
- Detection of multiple pathways in the spinal cord using q-ball imagingNeuroImage, 2008
- A 128‐channel receive‐only cardiac coil for highly accelerated cardiac MRI at 3 TeslaMagnetic Resonance in Medicine, 2008
- In vivo DTI of the healthy and injured cat spinal cord at high spatial and angular resolutionNeuroImage, 2008
- Brain morphometry with multiecho MPRAGENeuroImage, 2008
- Transmit and receive transmission line arrays for 7 Tesla parallel imagingMagnetic Resonance in Medicine, 2005
- Generalized autocalibrating partially parallel acquisitions (GRAPPA)Magnetic Resonance in Medicine, 2002