Force distributions associated with different elastic traction methods for maxillary dentition distalization by clear aligners: an in-vitro study
Authors:
Open Access
- 18 April 2025
- journal article
- research article
- Published by Springer Nature in BMC Oral Health
- Vol. 25 (1) , 1-10
Abstract
The aim of this study was to examine the distribution patterns of orthodontic and reactive forces on each tooth during dentition distalization using clear aligners (CAs), and to assess the impact of different elastic traction methods. Three sets of aligners for maxillary dentition distalization were fabricated, targeting different tooth movements: simultaneous distalization of the first molars and second premolars; simultaneous distalization of the first premolars and canines; and simultaneous retraction of incisors. An in vitro orthodontic simulator, consisting of 14 three-dimensional sensors, was used to evaluate mechanical changes in each tooth. The orthodontic and reactive forces exerted by aligners, along with additional pulling forces from 1/4 inch 3.5oz elastic using different elastic traction methods (with hooks or buttons), were measured and subjected to comparative analysis. The initial orthodontic force exerted by CAs varied across different teeth, ranging from 1.52 N to 6.77 N. Concurrently, the primary anchorage teeth experience reactive forces within a range of 1 N to 6.48 N. While the traction force on these teeth generally remained significantly smaller, staying below 0.6 N. The traction force exerted via hooks decayed to 84.3% and was primarily concentrated on the canines, whereas traction force applied via buttons transmitted approximately 96.1% and was more readily distributed to other teeth. Elastic traction is not sufficient to completely counteract the initial reactive forces produced by the deformation of CAs. It is advisable to use stronger elastics or extend aligner wear time to ensure sufficient anchorage protection. Utilizing traction with buttons can reduce mechanical loss and enhance the transmission of traction force to other teeth.Keywords
This publication has 32 references indexed in Scilit:
- Accuracy of maxillary molar distalization with clear aligners in three-dimension: a retrospective study based on CBCT superimpositionClinical Oral Investigations, 2025
- Fused filament fabrication (FFF): influence of layer height on forces and moments delivered by aligners—an in vitro studyClinical Oral Investigations, 2023
- G-Block: Posterior anchorage device tads-supported after molar distalization with aligners: An adult case reportInternational Orthodontics, 2022
- Force delivery modification of removable thermoplastic appliances using Hilliard precision thermopliers for tipping an upper central incisorClinical Oral Investigations, 2022
- An in vitro evaluation of orthodontic aligner biomechanics around the maxillary archAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2021
- Force degradation of orthodontic latex elastics analyzed in vivo and in vitroAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2020
- Effect of in vitro aging by water immersion and thermocycling on the mechanical properties of PETG aligner materialJournal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie, 2019
- Forces and moments applied during derotation of a maxillary central incisor with thinner aligners: An in-vitro studyAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2017
- Forces and moments delivered by PET-G aligners to an upper central incisor for labial and palatal translationJournal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie, 2015
- Forces and moments generated by removable thermoplastic aligners: Incisor torque, premolar derotation, and molar distalizationAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2014