Medical or Surgical Complications Related to the Nucleus Multichannel Cochlear Implant
Authors:
- 1 September 1988
- journal article
- research article
- Published by SAGE Publications in Annals of Otology, Rhinology & Laryngology
- Vol. 97 (5_suppl2) , 8-13
Abstract
A questionnaire was sent to 152 surgeons to survey complications associated with the implantation of the Nucleus multichannel cochlear implant. Complications were categorized as life-threatening; major, if they necessitated revision surgery; or minor, if they resolved spontaneously or with minimal treatment. A total of 55 complications occurred in 459 reported operations for an overall complication rate of 11.8%. There were no deaths, but there was one life-threatening complication, a case of meningitis. There were 23 (4.8%) major complications, most of which involved flap design or electrode insertion (and included the case of meningitis). There were 32 (7 %) minor complications. Most of the complications might have been avoided by proper training, planning of the operations, and careful attention to detail. We recommend that all prospective implant surgeons attend a device-specific training course and practice in the temporal bone laboratory.Keywords
This publication has 6 references indexed in Scilit:
- Flap design for cochlear implantation: Avoidance of a potential complicationsThe Laryngoscope, 1987
- Surgery for the Safe Insertion and Reinsertion of the Banded Electrode ArrayAnnals of Otology, Rhinology & Laryngology, 1987
- Problems and Complications of Cochlear Implant SurgeryAnnals of Otology, Rhinology & Laryngology, 1987
- House 3M Cochlear Implant: Surgical ConsiderationsAnnals of Otology, Rhinology & Laryngology, 1987
- Safety Considerations of Cochlear ImplantationOtolaryngologic Clinics of North America, 1986
- Banded Intracochlear Electrode Array: Evaluation of Insertion Trauma in Human Temporal BonesAnnals of Otology, Rhinology & Laryngology, 1985