Development of endovascular stent-grafts
Authors:
- 1 June 2007
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 221 (6) , 547-560
Abstract
The very first experiences in the early 1990s with endovascular aortic stent-grafts were associated with significant numbers of complications including an inability to deploy the stent-graft, conversion to open surgery, and aneurysm rupture. By the mid-1990s, improved home-made and commercially available stent-grafts started to appear. These devices could be successfully deployed in the aorta, achieving aneurysm exclusion with low morbidity and mortality. However, follow-up results raised concerns about the longer-term durability. Gradually, too, these problems have been addressed such that, in the recent UK multi-centre randomized controlled trial of endovascular versus open aneurysm repair, aneurysm-related mortality was 3 per cent less in the endovascular group four years following surgery. Currently the indications for aortic stent-grafts are being expanded. It is now possible to maintain perfusion successfully in aortic side branches and to treat aneurysms that would have once been thought untreatable. This review paper reviews the main developments in endovascular stent-grafting and the major role played by medical engineering and technology.Keywords
This publication has 58 references indexed in Scilit:
- Effect of radioactivity on stent-graft incorporation after endovascular treatment of aneurysms: An animal studyJournal of Biomedical Materials Research Part A, 2006
- In Vitro Analysis of Modular Aortic Stent-Graft FailureJournal of Endovascular Therapy, 2006
- A mathematical model to predict the in vivo pulsatile drag forces acting on bifurcated stent grafts used in endovascular treatment of abdominal aortic aneurysms (AAA)Journal of Biomechanics, 2004
- Explant analysis of AneuRx stent grafts: relationship between structural findings and clinical outcomeJournal of Vascular Surgery, 2004
- Management of Aneurysm Sac HygromaJournal of Endovascular Therapy, 2004
- Ingrowth of aorta wall into stent grafts impregnated with basic fibroblast growth factor: a porcine in vivo study of blood vessel prosthesis healingJournal of Vascular Surgery, 2004
- Fixation of Infrarenal Aortic Stent-grafts Using Laparoscopic Banding – An Experimental Study in PigsEuropean Journal of Vascular and Endovascular Surgery, 2001
- Aneurysm pressure following endovascular exclusionEuropean Journal of Vascular and Endovascular Surgery, 1997
- Invivo characterization of a fluoropassivated gelatin-impregnated polyester mesh for hernia repairJournal of Biomedical Materials Research, 1996