Resolution characterization of a silicon-based, photon-counting computed tomography prototype capable of patient scanning
- 15 October 2019
- journal article
- research article
- Published by SPIE-Intl Soc Optical Eng in Journal of Medical Imaging
- Vol. 6 (04) , 043502-9
Abstract
Photon-counting detectors are expected to bring a range of improvements to patient imaging with x-ray computed tomography (CT). One is higher spatial resolution. We demonstrate the resolution obtained using a commercial CT scanner where the original energy-integrating detector has been replaced by a single-slice, silicon-based, photon-counting detector. This prototype constitutes the first full-field-of-view silicon-based CT scanner capable of patient scanning. First, the pixel response function and focal spot profile are measured and, combining the two, the system modulation transfer function is calculated. Second, the prototype is used to scan a resolution phantom and a skull phantom. The resolution images are compared to images from a state-of-the-art CT scanner. The comparison shows that for the prototype 19 lp / cm are detectable with the same clarity as 14 lp / cm on the reference scanner at equal dose and reconstruction grid, with more line pairs visible with increasing dose and decreasing image pixel size. The high spatial resolution remains evident in the anatomy of the skull phantom and is comparable to that of other photon-counting CT prototypes present in the literature. We conclude that the deep silicon-based detector used in our study could provide improved spatial resolution in patient imaging without increasing the x-ray dose.Keywords
Funding Information
- European Union Horizon 2020 (750564)
This publication has 44 references indexed in Scilit:
- Investigating the effect of characteristic x-rays in cadmium zinc telluride detectors under breast computerized tomography operating conditionsJournal of Applied Physics, 2013
- Vision 20/20: Single photon counting x‐ray detectors in medical imagingMedical Physics, 2013
- To bin or not to bin? The effect of CT system limiting resolution on noise and detectabilityPhysics in Medicine & Biology, 2013
- Imaging properties of small-pixel spectroscopic x-ray detectors based on cadmium telluride sensorsPhysics in Medicine & Biology, 2012
- Photon counting multienergy x‐ray imaging: Effect of the characteristic x rays on detector performanceMedical Physics, 2009
- Optimal “image‐based” weighting for energy‐resolved CTMedical Physics, 2009
- Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomographyPhysics in Medicine & Biology, 2008
- Recent progress in CdTe and CdZnTe detectorsIEEE Transactions on Nuclear Science, 2001
- A method for measuring the presampled MTF of digital radiographic systems using an edge test deviceMedical Physics, 1998
- Energy-selective reconstructions in X-ray computerised tomographyPhysics in Medicine & Biology, 1976