Rapid T1 mapping of mouse myocardium with saturation recovery look‐locker method
Authors:
Open Access
- 26 October 2010
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 64 (5) , 1296-1303
Abstract
Dynamic contrast‐enhanced MRI using gadolinium or manganese provides unique characterization of myocardium and its pathology. In this study, an electrocardiography (ECG) triggered saturation recovery Look‐Locker method was developed and validated for fast cardiac T1 mapping in small animal models. By sampling the initial portion of the longitudinal magnetization recovery curve, high temporal resolution (∼3 min) can be achieved at a high spatial resolution (195 × 390 μm2) in mouse heart without the aid of parallel imaging or echo‐planar imaging. Validation studies were performed both in vitro on a phantom and in vivo on C57BL/6 mice (n = 6). Our results showed a strong agreement between T1 measured by saturation recovery Look‐Locker and by the standard saturation recovery method in vitro or inversion recovery Look‐Locker in vivo. The utility of saturation recovery Look‐Locker in dynamic contrast‐enhanced MRI studies was demonstrated in manganese‐enhanced MRI experiments in mice. Our results suggest that saturation recovery Look‐Locker can provide rapid and accurate cardiac T1 mapping for studies using small animal models. Magn Reson Med, 2010.Keywords
Funding Information
- NIH (HL73315, HL86935)
- Case Center for Imaging Research (U24 CA110943)
This publication has 33 references indexed in Scilit:
- Transmural myocardial strain in mouse: Quantification of high‐resolution MR tagging using harmonic phase (HARP) analysisMagnetic Resonance in Medicine, 2009
- An apparent unidirectional influx constant for manganese as a measure of myocardial calcium channel activityJournal of Magnetic Resonance Imaging, 2006
- T1 Measurements incorporating flip angle calibration and correction in vivoJournal of Magnetic Resonance, 2006
- Saturated double‐angle method for rapid B1+ mappingMagnetic Resonance in Medicine, 2006
- Fast high‐resolution T1 mapping of the human brainMagnetic Resonance in Medicine, 2005
- Comparative study of methods for determining vascular permeability and blood volume in human gliomasJournal of Magnetic Resonance Imaging, 2004
- Modified Look‐Locker inversion recovery (MOLLI) for high‐resolution T1 mapping of the heartMagnetic Resonance in Medicine, 2004
- Quantification of T1 values by SNAPSHOT-FLASH NMR imagingJournal of Magnetic Resonance (1969), 1992
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Variations in slice shape and absorption as artifacts in the determination of tissue parameters in NMR imagingMagnetic Resonance in Medicine, 1985