Characteristics of novel Ti–10Mo-xCu alloy by powder metallurgy for potential biomedical implant applications
Open Access
- 1 September 2020
- journal article
- research article
- Published by KEAI Publishing in Bioactive Materials
- Vol. 5 (3) , 659-666
Abstract
When biomaterials are implanted in the human body, the surfaces of the implants become favorable sites for microbial adhesion and biofilm formation, causing peri-implant infection which frequently results in the failure of prosthetics and revision surgery. Ti-Mo alloy is one of the commonly used implant materials for load-bearing bone replacement, and the prevention of infection of Ti-Mo implants is therefore crucial. In this study, bacterial inhibitory copper (Cu) was added to Ti-Mo matrix to develop a novel Ti-Mo-Cu alloy with bacterial inhibitory property. The effects of Cu content on microstructure, tensile properties, cytocompatibility, and bacterial inhibitory ability of Ti-Mo-Cu alloy were systematically investigated. Results revealed that Ti-10Mo-1Cu alloy consisted of α and β phases, while there were a few Ti2Cu intermetallic compounds existed for Ti-10Mo-3Cu and Ti-10Mo-5Cu alloys, in addition to α and β phases. The tensile strength of Ti-10Mo-xCu alloy increased with Cu content while elongation decreased. Ti-10Mo-3Cu alloy exhibited an optimal tensile strength of 1098.1 MPa and elongation of 5.2%. Cytocompatibility study indicated that none of the Ti-10Mo-xCu alloys had a negative effect on MC3T3-E1 cell proliferation. Bacterial inhibitory rates against S. aureus and E. coli increased with the increase in Cu content of Ti-10Mo-xCu alloy, within the ranges of 20-60% and 15-50%, respectively. Taken together, this study suggests that Ti-10Mo-3Cu alloy with high strength, acceptable elongation, excellent cytocompatibility, and the bacterial inhibitory property is a promising candidate for biomedical implant applications.Keywords
Funding Information
- Fundamental Research Funds for the Central Universities (FRF-TP-19005C1Z)
- China Scholarship Council (201906460106)
- Mauritius Research Council (MR/S010343/1)
- National Natural Science Foundation of China (51672184, 51874037, 51922004)
- European Commission
- University of Science and Technology Beijing (2019-Z14)
This publication has 36 references indexed in Scilit:
- Antibacterial effect of 317L stainless steel contained copper in prevention of implant-related infection in vitro and in vivoJournal of Materials Science: Materials in Medicine, 2011
- Prevention of pin tract infection with titanium‐copper alloysJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2009
- Ti based biomaterials, the ultimate choice for orthopaedic implants – A reviewProgress in Materials Science, 2009
- Electrochemical stability and corrosion resistance of Ti–Mo alloys for biomedical applicationsActa Biomaterialia, 2009
- Modification of medical metals by ion implantation of copperApplied Surface Science, 2007
- A novel antibacterial titania coating: Metal ion toxicity and in vitro surface colonizationJournal of Materials Science: Materials in Medicine, 2005
- Mechanical properties of cast Ti‐6Al‐4V‐XCu alloysJournal of Oral Rehabilitation, 2004
- Infections Associated with Surgical ImplantsNew England Journal of Medicine, 2004
- Metallurgical failure analysis of various implant materials used in orthopedic applicationsJournal of Failure Analysis and Prevention, 2004