In high performance liquid chromatographic methods using spectrophotometric detectors, a simple way to check the peak purity or homogeneity of a component is by the use of a signal (absorbance) ratio method. However, certain software limitations in the detectors exist. Using valerophenone as a sample component, a study was made of the data collected with a multiwavelength monitoring spectrophotometric detector. With the original software, the scale for the absorbance ratio axis was selected automatically and varied with each plot, giving all plots a similar magnitude. Misleading information on peak homogeneity was obtained when the absorbance ratio is close to one. This problem and others may be avoided through modifications of the subroutines in the original software.