Theoretical model of the response of pH‐based enzyme sensors
Authors:
- 1 May 1991
- journal article
- Published by Wiley in Electroanalysis
- Vol. 3 (4-5) , 361-364
Abstract
A theoretical kinetic model of a pH‐based enzymatic sensor has been developed. The transport phenomena of substrates and products of enzymatic reactions as well as of buffer components were taken into account. Protolytic equilibria were described by dissociation constants. It was also assumed that the kinetics of the enzymatic reaction is described by the Michaelis–Menten equation. The final equation for the steady‐state response of an enzyme‐based pH sensor describes the form of the calibration curve for this type of biosensor. This equation takes into account sample pH and buffer concentration, as well as kinetic parameters (Vmax, KM) of the enzymatic reaction. The proposed model is particularly useful for describing pH‐based enzyme sensors with thin enzymatic membranes.Keywords
This publication has 8 references indexed in Scilit:
- An algebraic equation for the steady-state response of enzyme-pH electrodes and field effect transistorsBiosensors, 1988
- Design of enzyme‐pH electrodesAIChE Journal, 1987
- pH-based enzyme potentiometric sensors. Part 1. TheoryAnalytical Chemistry, 1985
- Response of an enzyme-modified pH-sensitive ion selective device; consideration of the influence of the buffering capacity of the analyte solutionSensors and Actuators, 1985
- Steady-state theory of biocatalytic membrane electrodesAnalytical Chemistry, 1981
- Theoretical evaluation of the performance of enzyme electrodes and of enzyme reactorsMicrochimica Acta, 1980
- Kinetic behavior of enzymes immobilized in artificial membranesAnalytical Chemistry, 1972
- Some aspects of the kinetics of enzymic reactionsBiochimica et Biophysica Acta, 1953