Vibrational Magnetic Field Distortion Compensation System in MRI Superconducting Magnets
- 30 August 2013
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Applied Superconductivity
- Vol. 24 (3) , 1-4
Abstract
Superconducting magnets in MRIs are subject to various sources of vibrations. As the magnets respond to these excitations, eddy currents are generated in metallic structures in the cryostat and create their own magnetic fields inside the region of interest (ROI) where the imaging process takes place. Because these vibration-induced magnetic field distortions are not synchronized with pulse sequences, they are especially detrimental to image quality. In this paper, a novel passive compensation method for vibrational magnetic field distortions is presented. It provides a cost-effective solution that is capable of compensating dynamic field distortions over broad-band frequency ranges and complex vibrational behaviors.Keywords
This publication has 5 references indexed in Scilit:
- Phase errors in FSE signals due to low frequency electromagnetic interferenceMagnetic Resonance Imaging, 2013
- Novel technologies and configurations of superconducting magnets for MRISuperconductor Science and Technology, 2013
- Environmental Vibration Induced Magnetic Field Disturbance in MRI MagnetIEEE Transactions on Applied Superconductivity, 2011
- Correction for eddy current induced Bo shifts in diffusion-weighted echo-planar imagingMagnetic Resonance in Medicine, 1999
- Characterization of and correction for eddy current artifacts in echo planar diffusion imagingMagnetic Resonance in Medicine, 1998