At DT-fueled fusion reactors, there will be a need for tritium monitors that can simultaneously measure in real time the concentrations of HTO, HT and the activated air produced by fusion neutrons. Such a monitor has been developed, tested and delivered to the Princeton Plasma Physics Laboratory for use at the Tokamak Fusion Test Reactor (TFTR). It uses semipermeable membranes to achieve the removal of HTO from the sampled air for monitoring and a catalyst to convert the HT to HTO, also for removal and monitoring. The remaining air, devoid of tritium, is routed to a third detector for monitoring the activated air. Ihe sensitivities are those that would be expected fron tritium instruments employing conventional flow-through Ionization chambers: 1–3 µCi/m3. Its discriminating ability is approximately 10−3 for any of the three components (HTO, HT and activated air) in any of the other two channels. For instance, the concentration of HT in the HTO channel is 10−3 times its original concentration in t...Keywords
This publication has 5 references indexed in Scilit:
- Tritium Monitoring Within the Reactor Hall of a DT Fusion ReactorNuclear Technology - Fusion, 1983
- Tritium monitor for fusion reactors1982
- Nuclear radiation analysis for TFTR1980
- Separation of HT, noble gases and HTO vapor with semipermeable membranesNuclear Instruments and Methods, 1975
- Automated Nondestructive Assay Instrumentation for Nuclear Materials SafeguardsIEEE Transactions on Nuclear Science, 1975