Clinical Accuracy of a Continuous Glucose Monitoring System With an Advanced Algorithm
Open Access
- 3 November 2014
- journal article
- research article
- Published by SAGE Publications in Journal of Diabetes Science and Technology
- Vol. 9 (2) , 209-214
Abstract
Background: We assessed the performance of a modified Dexcom G4 Platinum system with an advanced algorithm, in comparison with frequent venous samples measured on a laboratory reference (YSI) during a clinic session and in comparison to self-monitored blood glucose (SMBG) during home use. Methods: Fifty-one subjects with diabetes were enrolled in a prospective multicenter study. Subjects wore 1 sensor for 7-day use and participated in one 12-hour in-clinic session on day 1, 4, or 7 to collect YSI reference venous glucose every 15 minutes and capillary SMBG test every 30 minutes. Carbohydrate consumption and insulin dosing and timing were manipulated to obtain data in low and high glucose ranges. Results: In comparison with the laboratory reference method (n = 2,263) the system provided a mean and median absolute relative differences (ARD) of 9.0% and 7.0%, respectively. The mean absolute difference for CGM was 6.4 mg/dL when the YSIs were within hypoglycemia ranges (≤ 70 mg/dL). The percentage in the clinically accurate Clarke error grid A zone was 92.4% and in the benign error B zone was 7.1%. Majority of the sensors (73%) had an aggregated MARD in reference to YSI ≤ 10%. The MARD of CGM-SMBG for home use was 11.3%. Conclusions: The study showed that the point and rate accuracy, clinical accuracy, reliability, and consistency over the duration of wear and across glycemic ranges were superior to current commercial real-time CGM systems. The performance of this CGM is reaching that of a self-monitoring blood glucose meter in real use environment.Keywords
This publication has 21 references indexed in Scilit:
- A Clinical Trial of the Accuracy and Treatment Experience of the Dexcom G4 Sensor (Dexcom G4 System) and Enlite Sensor (Guardian REAL-Time System) Tested Simultaneously in Ambulatory Patients with Type 1 DiabetesDiabetes Technology & Therapeutics, 2014
- Accuracy of two continuous glucose monitoring systems: a head‐to‐head comparison under clinical research centre and daily life conditionsDiabetes, Obesity and Metabolism, 2014
- Accuracy and Acceptability of the 6-Day Enlite Continuous Subcutaneous Glucose SensorDiabetes Technology & Therapeutics, 2014
- A Comparative Effectiveness Analysis of Three Continuous Glucose MonitorsJournal of Diabetes Science and Technology, 2014
- Nocturnal Continuous Glucose Monitoring: Accuracy and Reliability of Hypoglycemia Detection in Patients with Type 1 Diabetes at High Risk of Severe HypoglycemiaDiabetes Technology & Therapeutics, 2013
- Performance Variability of Seven Commonly Used Self-Monitoring of Blood Glucose Systems: Clinical Considerations for Patients and ProvidersJournal of Diabetes Science and Technology, 2013
- (Analytical) Accuracy of Blood Glucose Meters and Patients: How Do They Come Together?Journal of Diabetes Science and Technology, 2013
- System Accuracy Evaluation of 43 Blood Glucose Monitoring Systems for Self-Monitoring of Blood Glucose according to DIN EN ISO 15197Journal of Diabetes Science and Technology, 2012
- How to Assess and Compare the Accuracy of Continuous Glucose Monitors?Diabetes Technology & Therapeutics, 2008
- Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood GlucoseDiabetes Care, 1987