Dataset of manually measured QT intervals in the electrocardiogram
Open Access
- 18 May 2006
- journal article
- Published by Springer Nature in BioMedical Engineering OnLine
- Vol. 5 (1) , 31
Abstract
The QT interval and the QT dispersion are currently a subject of considerable interest. Cardiac repolarization delay is known to favor the development of arrhythmias. The QT dispersion, defined as the difference between the longest and the shortest QT intervals or as the standard deviation of the QT duration in the 12-lead ECG is assumed to be reliable predictor of cardiovascular mortality. The seventh annual PhysioNet/Computers in Cardiology Challenge, 2006 addresses a question of high clinical interest: Can the QT interval be measured by fully automated methods with accuracy acceptable for clinical evaluations? The PTB Diagnostic ECG Database was given to 4 cardiologists and 1 biomedical engineer for manual marking of QRS onsets and T-wave ends in 458 recordings. Each recording consisted of one selected beat in lead II, chosen visually to have minimum baseline shift, noise, and artifact.In cases where no T wave could be observed or its amplitude was very small, the referees were instructed to mark a 'group-T-wave end' taking into consideration leads with better manifested T wave.A modified Delphi approach was used, which included up to three rounds of measurements to obtain results closer to the median. A total amount of 2*5*548 Q-onsets and T-wave ends were manually marked during round 1. To obtain closer to the median results, 8.58 % of Q-onsets and 3.21 % of the T-wave ends had to be reviewed during round 2, and 1.50 % Q-onsets and 1.17 % T-wave ends in round 3. The mean and standard deviation of the differences between the values of the referees and the median after round 3 were 2.43 +/- 0.96 ms for the Q-onset, and 7.43 +/- 3.44 ms for the T-wave end. A fully accessible, on the Internet, dataset of manually measured Q-onsets and T-wave ends was created and presented in additional file: 1 (Table 4) with this article. Thus, an available standard can be used for the development of automated methods for the detection of Q-onsets, T-wave ends and for QT interval measurements.Keywords
This publication has 18 references indexed in Scilit:
- A Wavelet-Based ECG Delineator: Evaluation on Standard DatabasesIEEE Transactions on Biomedical Engineering, 2004
- Computerized QT dispersion measurement and cardiovascular mortality in male veteransThe American Journal of Cardiology, 2004
- Interrater Agreement in the Measurement of QRS Interval in Tricyclic Antidepressant Overdose: Implications for Monitoring and ResearchAnnals of Emergency Medicine, 1996
- Commentary: QTc dispersion may reflect vulnerability to ventricular fibrillationBMJ, 1996
- Variation in the identification of Q wave initiation and its contribution to QT measurementPhysiological Measurement, 1995
- Errors in manual measurement of QT intervals.Heart, 1994
- Multilead ECG analysisComputers and Biomedical Research, 1986
- Assessment of the performance of electrocardiographic computer programs with the use of a reference data base.Circulation, 1985
- Errors in the visual determination of corrected QT (QTc) interval during acute myocardial infarctionJACC, 1985
- An experimental study of group opinionFutures, 1969