Dual Energy Analysis Using Phoswich Scintillation Detectors for Low-Level In-Vivo Counting
Authors:
- 1 February 1974
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 21 (1) , 201-209
Abstract
Electronics developed for a whole body counting system analyze the pulse heights of signals from both the sodium iodide (NaI(Tl)) and cesium iodide (CsI(Na)) crystals of a phoswich scintillation detector. Pulse shape discrimination is used to determine the origin of the scintillation, NaI or CsI, while rejecting coincident Compton-scattered events and photomultiplier tube noise. Both energy spectra are accumulated simultaneously in separate memory quadrants of a multichannel analyzer. The system allows for low-energy (5 keV - 125 keV) x-ray analysis from the NaI(Tl) crystal and high-energy (100 keV - 2.5 MeV) x-ray or γ-ray analysis from the CsI(Na) crystal. The discrimination system is better than 95% efficient over both energy ranges, with a low-energy (12-24keV) shielded background count of less than 2 cpm for a single 5-in.-diam phoswich detector. This system thus extends the detector's normal energy range by more than an order of magnitude.Keywords
This publication has 6 references indexed in Scilit:
- Pulse Shape Discriminators for NaI - CsI Phoswich DetectorsIEEE Transactions on Nuclear Science, 1972
- A Pulse Shape Analyzer for Phoswich DetectorsIEEE Transactions on Nuclear Science, 1972
- On the construction of a fast constant fraction trigger with integrated circuits and application to various photomultipliertubesNuclear Instruments and Methods, 1970
- A general purpose pulse shape discriminating circuitNuclear Instruments and Methods, 1969
- Time resolution studies on large photomultipliersNuclear Instruments and Methods, 1967
- The application of pulse-shape discrimanation to separating phototube noise pulse from scintillation pulsesNuclear Instruments and Methods, 1965