Bias Calibration of Optically Pumped Magnetometers Based on Variable Sensitivity
Open Access
Abstract
Optically pumped magnetometers (OPMs) functioning in the spin-exchange relaxation-free (SERF) regime have emerged as attractive options for measuring weak magnetic fields, owing to their portability and remarkable sensitivity. The operation of SERF-OPM critically relies on the ambient magnetic field; thus, a magnetic field compensation device is commonly employed to mitigate the ambient magnetic field to near zero. Nonetheless, the bias of the OPM may influence the compensation impact, a subject that remains unexamined. This paper introduced an innovative bias calibration technique for OPMs. The sensitivity of the OPM was altered by adjusting the cell temperature. The output of the OPM was then documented with varying sensitivity. It is assumed that the signal exhibits a linear correlation with the environmental magnetic field, and the statistical characteristics of the magnetic field are identical for both measurements, upon which the bias of the OPM is assessed. The bias was subsequently considered in the feedback magnetic field compensation mechanism. The results indicate that this technique might successfully reduce environmental magnetic fluctuations and enhance the sensitivity of the OPM. This technique measured the magnetic field produced by the human heart, confirming the viability of the ultra-weak biomagnetic field measurement approach.Keywords
Funding Information
- National Key R&D Program of China: BTIT (2022YFF1202800)
This publication has 39 references indexed in Scilit:
- Optical pump magnetometers parametric correction method based on three-axis coil arraysMeasurement, 2024
- Analysis and suppression of cross-axis modulation coupling errors in optically pumped magnetometersMeasurement, 2024
- Fast In-Situ Triaxial Remanent Magnetic Field Measurement for Single-Beam SERF Atomic Magnetometer Based on Trisection AlgorithmPhotonic Sensors, 2023
- Temperature characteristics of Rb-N2 single-beam magnetometer with different buffer gas pressuresMeasurement, 2023
- Measuring functional connectivity with wearable MEGNeuroImage, 2021
- Noninvasive measurement of sensory action currents in the cervical cord by magnetospinographyClinical Neurophysiology, 2021
- Optically pumped magnetometers disclose magnetic field components of the muscular action potentialJournal of Electromyography and Kinesiology, 2021
- Mouth magnetoencephalography: A unique perspective on the human hippocampusNeuroImage, 2021
- Optically pumped magnetometers: From quantum origins to multi-channel magnetoencephalographyNeuroImage, 2019
- A bi-planar coil system for nulling background magnetic fields in scalp mounted magnetoencephalographyNeuroImage, 2018