Active matrix liquid crystal displays for clinical imaging: Comparison with cathode ray tube displays
- 1 May 2000
- journal article
- research article
- Published by Springer Nature in Journal of Digital Imaging
- Vol. 13 (1) , 155-161
Abstract
Fifteen large-area, flat-panel displays used for clinical image review were evaluated for image quality and compared with 30 comparably sized cathode ray tube (CRT) monitors. Measurements were of image display patterns by Video Electronic Standards Association (VESA) and a commercial product. Field measurements were made of maximum and minimum luminance, ambient lighting, characteristic curve (gamma), point shape and size, high-contrast resolution, uniformity, and distortion. Assessments were made of pixel defects, latent image patterns, ghosting artifacts, and viewing angle luminance. Also, a questionnaire was generated for users of the flat-panel and CRT units. Seventeen respondents indicated no preference for either flat panel or CRT. Results show these flat panels to have higher luminance (mean, 177.7 cd/m2); larger number of just noticeable differences (JNDs; n=555), higher gamma, comparable uniformity, and warm-up time. CRTs had less angle viewing dependence and far fewer artifacts (ghosting and latent images). Our questionnaire showed active matrix liquid crystal displays (AMLCD) to be fully acceptable for clinical image viewing. Furthermore, the statistical results show that further testing for new AMLCDs of this type is unwarranted.Keywords
This publication has 19 references indexed in Scilit:
- Characterization of color CRT display systems for monochrome applicationsJournal of Digital Imaging, 1999
- Image quality degradation by light scattering in display devicesJournal of Digital Imaging, 1999
- Theoretical and experimental study of nematic liquid crystal display cells using the in-plane-switching modeIEEE Transactions on Electron Devices, 1999
- CRT versus LCD: A pilot study on visual performance and suitability of two display technologies for use in office workDisplays, 1999
- Guest editorial: Image quality assurance for CRT display systemsJournal of Digital Imaging, 1999
- Glare from monitors measured with subjective scales and eye movementsDisplays, 1999
- Specular and diffuse reflection measurement feasibility study of ISO9241 Part 7 methodDisplays, 1998
- Statistical methods for testing the visual quality of displaysDisplays, 1997
- The optimal glass transmission coefficient as a function of the level of diffuse ambient illuminationDisplays, 1997
- Sunlight viewable electroluminescent displays for military applicationIEEE Aerospace and Electronic Systems Magazine, 1995