Development and Validation of a Photoplethysmography System for Noninvasive Monitoring of Hemoglobin Concentration
Authors:
Open Access
- 18 January 2020
- journal article
- research article
- Published by Wiley in Journal of Electrical and Computer Engineering
- Vol. 2020, 1-9
Abstract
Background. Traditional invasive hemoglobin (Hb) detection led to delayed diagnosis, operational inefficiency, incorrect critical decision making, and uncomfortable patient experience. To facilitate real-time total hemoglobin (tHb) monitoring, a portable prototype of a noninvasive Hb detection system was developed, and the accuracy of Hb predicted based on partial least squares (PLS) as well as backpropagation artificial neural network (BP-ANN) models was validated. Results. The prototype was combined with a signal processing circuit and a spectrophotometric probe containing 8 wavelength LEDs light source and photodiode array. Laboratory invasive Hb (Lab_tHb) and spot check Hb measurements with PLS (SpHb_PLS) and BP (SpHb_BP) methods were obtained simultaneously by hematology analyzer and the designed system. The invasive and noninvasive estimates of the Hb levels were analyzed using Spearman correlation as well as Bland–Altman plot and receiver operating characteristic (ROC) curve analysis. A total of 238 volunteers had attempted laboratory invasive and noninvasive spot check Hb measurements. Mean Lab_tHb, SpHb_PLS, and SpHb_BP were 13.6 ± 1.80 g/dL, 13.5 ± 1.07 g/dL, and 13.6 ± 1.06 g/dL, respectively. Noninvasive SpHb_PLS (r = 0.61, ) and SpHb_BP (r = 0.62, ) had a strong correlation with invasive tHb values. The Bland–Altman plot showed excellent consistency between the proposed noninvasive methods and laboratory invasive reference. In ROC analysis, PLS and BP models were good at predicting Hb ≥ 12 g/dL with area under the curve of 0.828 and 0.824, respectively. Observed differences between invasive and noninvasive Hb measurements displayed no significant correlation with perfusion index values. Conclusions. The result confirmed that noninvasive Hb monitoring had an excellent correlation with traditional invasive Hb measurement. Furthermore, it is suggested that the developed prototype has the potential for the noninvasive detection of Hb concentration with the methods of PLS and BP-ANN.
Keywords
Funding Information
- National Key Technology R&D Program (2016YFC1305703, 2013BAI03B03, 2015BAI01B08, 2016-tm-042, 61372047, 61701540)
- Ministry of Science and Technology of the People's Republic of China (2016YFC1305703, 2013BAI03B03, 2015BAI01B08, 2016-tm-042, 61372047, 61701540)
- Ministry of Science and Technology of the People's Republic of China (2016YFC1305703, 2013BAI03B03, 2015BAI01B08, 2016-tm-042, 61372047, 61701540)
- Ministry of Science and Technology of the People's Republic of China (2016YFC1305703, 2013BAI03B03, 2015BAI01B08, 2016-tm-042, 61372047, 61701540)
- Chinese PLA General Hospital (2016YFC1305703, 2013BAI03B03, 2015BAI01B08, 2016-tm-042, 61372047, 61701540)
- National Natural Science Foundation of China (2016YFC1305703, 2013BAI03B03, 2015BAI01B08, 2016-tm-042, 61372047, 61701540)
- National Natural Science Foundation of China (2016YFC1305703, 2013BAI03B03, 2015BAI01B08, 2016-tm-042, 61372047, 61701540)
This publication has 34 references indexed in Scilit:
- Non-Invasive Measurement of Hemoglobin: Assessment of Two Different Point-of-Care TechnologiesPLOS ONE, 2012
- A Comparison of Three Methods of Hemoglobin Monitoring in Patients Undergoing Spine SurgeryAnesthesia & Analgesia, 2011
- Data preprocessing and partial least squares regression analysis for reagentless determination of hemoglobin concentrations using conventional and total transmission spectroscopyJournal of Biomedical Optics, 2001
- Frequency-domain measurements of photon migration in tissuesChemical Physics Letters, 1990
- Time resolved reflectance and transmittance for the noninvasive measurement of tissue optical propertiesApplied Optics, 1989
- Multilayer feedforward networks are universal approximatorsNeural Networks, 1989
- Theory of the backpropagation neural networkNeural Networks, 1988
- Partial least-squares regression: a tutorialAnalytica Chimica Acta, 1986
- Wavelength dependency of the spectrophotometric determination of blood oxygen saturationClinica Chimica Acta, 1969
- International Committee for Standardization in HaematologyBritish Journal of Haematology, 1967