Evaluation of the Sysmex UF-1000i for the Diagnosis of Urinary Tract Infection
Authors:
Open Access
- 1 April 2010
- journal article
- research article
- Published by Oxford University Press (OUP) in American Journal of Clinical Pathology
- Vol. 133 (4) , 577-582
Abstract
Diagnosis of urinary tract infection (UTI) is primarily done by microbiologic culture, which is time-consuming and can produce false-positives and false-negatives. Flow cytometry allows for rapid screening of many samples and eliminates culturing. We analyzed the Sysmex UF-1000i (TOA Medical Electronics, Kobe, Japan) for accuracy in identifying RBCs and WBCs, casts, bacteria, and epithelia. We also evaluated its precision, linear estimation of results, carryover contamination rate, and anti-interference. UF-1000i agreement with manual counting was approximately 95% for RBCs and WBCs, epithelia, and casts. Its coefficient of variation for bacteria ranged from 4.7% to 15.2%. UF-1000i screening for UTIs exhibited great sensitivity (97%), specificity (79%), positive predictive value (70%), negative predictive value (99%), and accuracy (85%). The negative predictive value remained high even with complex UTI samples. The Sysmex UF-1000i shows great promise in excluding more than 50% of true-negative samples, improving detection efficiency, and reducing laboratory costs.Keywords
This publication has 16 references indexed in Scilit:
- Manufacturing fault causes a false negative for haematuria using Makromed m8l urinalysis test stripsMedicine, Science and the Law, 2009
- Diagnostic Accuracy of Criteria for Urinary Tract Infection in a Cohort of Nursing Home Residents *Journal of the American Geriatrics Society, 2007
- Diagnostic performance of combined specific urinary proteins and urinary flow cytometry in urinary tract pathologyClinical Chemistry and Laboratory Medicine (CCLM), 2007
- The prevalence of asymptomatic pyuria in diabetic patients: Comparison of the Sysmex UF-100 automated urinalysis analyzer with Fuchs–Rosenthal hemacytometerClinical Biochemistry, 2006
- Automated flow cytometry analysis of peritoneal dialysis fluidNephrology Dialysis Transplantation, 2004
- Unsatisfactory Performance of Flow Cytometer UF-100 and Urine Strips in Predicting Outcome of Urine CulturesJournal of Clinical Microbiology, 2001
- Automation of Urine Sediment Examination: a Comparison of the Sysmex UF-100 Automated Flow Cytometer with Routine Manual Diagnosis (Microscopy, Test Strips, and Bacterial Culture)Clinical Chemistry and Laboratory Medicine (CCLM), 1999
- The Automation of Sediment Urinalysis Using a New Urine Flow Cytometer (UF-100™)Clinical Chemistry and Laboratory Medicine (CCLM), 1998
- Bacterial Characterization by Flow CytometryScience, 1983
- On the methodology of investigations of etiologic factors in chronic diseasesJournal of Chronic Diseases, 1959