Nanoelectronic thermometers optimised for sub-10 millikelvin operation

Authors:
  • JPJ. R. Prance
    J. R. Prance
  • DBD. I. Bradley
    D. I. Bradley
  • RGR. E. George
    R. E. George
  • RHR. P. Haley
    R. P. Haley
  • YPYu. A. Pashkin
    Yu. A. Pashkin
  • MSM. Sarsby
    M. Sarsby
  • JPJ. Penttilä
    J. Penttilä
  • LRL. Roschier
    L. Roschier
  • DGD. Gunnarsson
    D. Gunnarsson
  • HHH. Heikkinen
    H. Heikkinen
  • MPM. Prunnila
    M. Prunnila

Abstract
We report the cooling of electrons in nanoelectronic Coulomb blockade thermometers below 4 mK. Above 7 mK the devices are in good thermal contact with the environment, well isolated from electrical noise, and not susceptible to self-heating. This is attributed to an optimised design that incorporates cooling fins with a high electron-phonon coupling and on-chip electronic filters, combined with a low-noise electronic measurement setup. Below 7 mK the electron temperature is seen to diverge from the ambient temperature. By immersing a Coulomb Blockade Thermometer in the 3He/4He refrigerant of a dilution refrigerator, we measure a lowest electron temperature of 3.7 mK and a trend to a saturated electron temperature of 1.0 mK.