Real-time vessel segmentation and reconstruction for virtual fixtures for an active handheld microneurosurgical instrument
- 16 March 2022
- journal article
- research article
- Published by Springer Nature in International Journal of Computer Assisted Radiology and Surgery
- Vol. 17 (6) , 1069-1077
Abstract
Purpose Complications related to vascular damage such as intra-operative bleeding may be avoided during neurosurgical procedures such as petroclival meningioma surgery. To address this and improve the patient’s safety, we designed a real-time blood vessel avoidance strategy that enables operation on deformable tissue during petroclival meningioma surgery using Micron, a handheld surgical robotic tool. Methods We integrated real-time intra-operative blood vessel segmentation of brain vasculature using deep learning, with a 3D reconstruction algorithm to obtain the vessel point cloud in real time. We then implemented a virtual-fixture-based strategy that prevented Micron’s tooltip from entering a forbidden region around the vessel, thus avoiding damage to it. Results We achieved a median Dice similarity coefficient of 0.97, 0.86, 0.87 and 0.77 on datasets of phantom blood vessels, petrosal vein, internal carotid artery and superficial vessels, respectively. We conducted trials with deformable clay vessel phantoms, keeping the forbidden region 400 \(\mu \)m outside and 400 \(\mu \)m inside the vessel. Micron’s tip entered the forbidden region with a median penetration of just 8.84 \(\mu \)m and 9.63 \(\mu \)m, compared to 148.74 \(\mu \)m and 117.17 \(\mu \)m without our strategy, for the former and latter trials, respectively. Conclusion Real-time control of Micron was achieved at 33.3 fps. We achieved improvements in real-time segmentation of brain vasculature from intra-operative images and showed that our approach works even on non-stationary vessel phantoms. The results suggest that by enabling precise, real-time control, we are one step closer to using Micron in real neurosurgical procedures.
Keywords
Funding Information
- National Institutes of Health (R01EB000526, R01EB024564)
This publication has 14 references indexed in Scilit:
- Effective Deep Learning for Semantic Segmentation Based Bleeding Zone Detection in Capsule Endoscopy Images2018
- Toward Improving Safety in Neurosurgery with an Active Handheld InstrumentAnnals of Biomedical Engineering, 2018
- Surgical tool segmentation using a hybrid deep CNN-RNN auto encoder-decoder2017
- Ultrastructural Characterization of the Lower Motor System in a Mouse Model of Krabbe DiseaseScientific Reports, 2016
- U-Net: Convolutional Networks for Biomedical Image Segmentation2015
- Preliminary evaluation of a micro-force sensing handheld robot for vitreoretinal surgery2012
- Micron: An Actively Stabilized Handheld Tool for MicrosurgeryIEEE Transactions on Robotics, 2011
- Incidence and predictors of major perioperative adverse cardiac and cerebrovascular events in non-cardiac surgeryBritish Journal of Anaesthesia, 2011
- High-Speed Microscale Optical Tracking Using Digital Frequency-Domain MultiplexingIEEE Transactions on Instrumentation and Measurement, 2008
- Petrosal Approach for Petroclival MeningiomasNeurosurgery, 1988