A manufacturer's role in reducing the dose of cone beam computed tomography examinations: effect of beam filtration
Author:
- 1 February 2011
- journal article
- research article
- Published by Oxford University Press (OUP) in Dentomaxillofacial Radiology
- Vol. 40 (2) , 115-122
Abstract
Objectives: The dosimetry of the Kodak 9500 cone beam CT (CBCT) unit (Carestream Health, Rochester, NY) was measured before and after installation of copper filtration. Methods: Dosimetry of a pre-production Kodak 9500 CBCT unit was compared with a current production unit with 0.4 mm of added filtration and increased kVp. Thermoluminescent dosimeter 100 chips were placed at 24 locations in a RANDO (radiation analogue dosimetry) head phantom (Nuclear Associates, Hicksville, NY). Small, medium and large adult default exposure settings were used in separate dosimeter runs for large and mediumfield of view (FOV) examinations with both units. Equivalent dose and effective dose were calculated using International Commission on Radiological Protection (ICRP) 1990 and 2007 tissue weights. Results: Estimations of risk using 2007 ICRP calculations increased by an average of 77% for large FOV scans and 125% for the medium FOV scans in comparison with 1990 calculations. With added filtration, effective dose for medium FOV examinations for default settings were: small adult 76 mu Sv, medium adult 98 mu Sv, and large adult 166 mSv. Effective doses for large FOV examinations were: small adult 93 mu Sv, medium adult 163 mu Sv, and large adult 260 mu Sv. Effective dose was reduced by an average of 43% in examinations made with increased filtration and adjusted kVp. Conclusion: The manufacturer's installation of additional filtration with the adjustment of kVp in the Kodak 9500 CBCT unit resulted in significant patient dose reductions for examinations at all adult default settings. Dentomaxillofacial Radiology (2011) 40, 115-122. doi: 10.1259/dmfr/31708191Keywords
This publication has 12 references indexed in Scilit:
- Radiation Dose Associated With Common Computed Tomography Examinations and the Associated Lifetime Attributable Risk of CancerArchives of internal medicine (1960), 2009
- Cone-beam computerized tomography (CBCT) imaging of the oral and maxillofacial region: A systematic review of the literatureInternational Journal of Oral and Maxillofacial Surgery, 2009
- Patient Risk Related to Common Dental Radiographic ExaminationsThe Journal of the American Dental Association, 2008
- Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiologyOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2008
- Computed Tomography — An Increasing Source of Radiation ExposureNew England Journal of Medicine, 2007
- Preface, Executive Summary and GlossaryAnnals of the ICRP, 2007
- Dosimetry of 3 CBCT devices for oral and maxillofacial radiology: CB Mercuray, NewTom 3G and i-CATDentomaxillofacial Radiology, 2006
- Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unitDentomaxillofacial Radiology, 2003
- IntroductionAnnals of the ICRP, 1991
- Radiobiologic risk estimation from dental radiologyOral Surgery, Oral Medicine, Oral Pathology, 1988