Accurate and precise calibration of AFM cantilever spring constants using laser Doppler vibrometry
Authors:
- Richard S GatesRGRichard S Gates
- Jon R PrattJPJon R Pratt
- 24 August 2012
- journal article
- Published by IOP Publishing in Nanotechnology
- Vol. 23 (37) , 375702
Abstract
Accurate cantilever spring constants are important in atomic force microscopy both in control of sensitive imaging and to provide correct nanomechanical property measurements. Conventional atomic force microscope (AFM) spring constant calibration techniques are usually performed in an AFM. They rely on significant handling and often require touching the cantilever probe tip to a surface to calibrate the optical lever sensitivity of the configuration. This can damage the tip. The thermal calibration technique developed for laser Doppler vibrometry (LDV) can be used to calibrate cantilevers without handling or touching the tip to a surface. Both flexural and torsional spring constants can be measured. Using both Euler-Bernoulli modeling and an SI traceable electrostatic force balance technique as a comparison we demonstrate that the LDV thermal technique is capable of providing rapid calibrations with a combination of ease, accuracy and precision beyond anything previously available.Keywords
This publication has 32 references indexed in Scilit:
- Characterization of application-specific probes for SPMs1997
- Determination of the spring constants of probes for force microscopy/spectroscopyNanotechnology, 1996
- A method for determining the spring constant of cantilevers for atomic force microscopyMeasurement Science and Technology, 1996
- Lateral, normal, and longitudinal spring constants of atomic force microscopy cantileversReview of Scientific Instruments, 1994
- Experimental Determination of Spring Constants in Atomic Force MicroscopyLangmuir, 1994
- A nondestructive method for determining the spring constant of cantilevers for scanning force microscopyReview of Scientific Instruments, 1993
- Micromachined silicon sensors for scanning force microscopyJournal of Vacuum Science & Technology B: Microelectronics and Nanometer Structures Processing, Measurement, and Phenomena, 1991
- Microfabrication of cantilever styli for the atomic force microscopeJournal of Vacuum Science & Technology A, 1990
- Atomic-scale friction of a tungsten tip on a graphite surfacePhysical Review Letters, 1987
- Atomic Force MicroscopePhysical Review Letters, 1986