Bioengineering reasons for the failure of metal-on-metal hip prostheses
Author:
- 1 August 2011
- journal article
- Published by British Editorial Society of Bone & Joint Surgery in The Journal of Bone and Joint Surgery. British volume
- Vol. 93-B (8) , 1001-1004
Abstract
Bioengineering reasons for increased wear and failure of metal-on-metal (MoM) bearings in hip prostheses have been described. Low wear occurs in MoM hips when the centre of the femoral head is concentric with the centre of the acetabular component and the implants are correctly positioned. Translational or rotational malpositioning of the components can lead to the contact-patch of the femoral component being displaced to the rim of the acetabular component, resulting in a ten- to 100-fold increase in wear and metal ion levels. This may cause adverse tissue reactions, loosening of components and failure of the prosthesis.Keywords
This publication has 24 references indexed in Scilit:
- Survival of Hard-on-Hard Bearings in Total Hip Arthroplasty: A Systematic ReviewClinical Orthopaedics and Related Research, 2010
- Effect of cup inclination angle during microseparation and rim loading on the wear of BIOLOX® delta ceramic‐on‐ceramic total hip replacementJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2010
- High Cup Angle and Microseparation Increase the Wear of Hip Surface ReplacementsClinical Orthopaedics and Related Research, 2009
- Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacementThe Journal of Bone and Joint Surgery. British volume, 2008
- Effect of bearing size on the long‐term wear, wear debris, and ion levels of large diameter metal‐on‐metal hip replacements—An in vitro studyJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2008
- Tribo-corrosion analysis of wear and metal ion release interactions from metal-on-metal and ceramic-on-metal contacts for the application in artificial hip prosthesesProceedings of the Institution of Mechanical Engineers, Part J: Journal of Engineering Tribology, 2008
- A comparison of friction in 28 mm conventional and 55 mm resurfacing metal-on-metal hip replacementsProceedings of the Institution of Mechanical Engineers, Part J: Journal of Engineering Tribology, 2007
- PRESIDENTIAL GUEST LECTURE: Tribology of Alternative BearingsClinical Orthopaedics and Related Research, 2006
- Friction of total hip replacements with different bearings and loading conditionsJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2006
- Metal‐on‐metal bearing wear with different swing phase loadsJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2004