Anesthesia depth monitoring using alternative placement of entropy sensors: a prospective study
- 19 December 2018
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 33 (5) , 871-876
Abstract
Spectral entropy is based on analysis of variations in electroencephalography and frontal electromyography, and is a safe and reliable method for anesthesia depth monitoring. However, standard frontal positioning of entropy electrodes in patients undergoing cardiac surgery is sometimes challenging. The present study aimed to compare standard entropy sensor placement with an alternative (infraorbital) site. This prospective study included 20 patients who underwent cardiovascular surgery at the authors’ center. Monitoring was performed with standard and alternative entropy electrode positions from patient admission to surgery to transfer to the intensive care unit. Data were recorded every 15 s; all data were analyzed and compared using Bland–Altman, scatter plot with Pearson correlation coefficient, and sensitivity/specificity analyses. Overall, 20,784 pairs of response entropy (RE) and state entropy (SE) indexes were collected. Bland–Altman analysis revealed a mean difference in RE of 0.37 (95% LOA − 7.09, 7.88) and SE 0.69 (95% LOA − 5.95, 7.31); with 3.46% (720/20,784) RE and 3.40% (706/10,790) SE values lying outside of the limits of agreement. Correlation analysis revealed strong positive correlation in both cases: RE, r = 0.983, p < 0.05; SE, r = 0.984; p < 0.05. Sensitivity/specificity analysis revealed 98.1% sensitivity, 93.3% specificity and 97.1% test efficiency for RE, and 99.2%, 95.1% and 98.5% for SE, respectively. Infraorbital entropy sensor placement in patients undergoing cardiovascular surgery is reliable and effective. The strong positive correlation between the two methods of registration enables alternative entropy measurement when frontal placement is not possible.Keywords
This publication has 12 references indexed in Scilit:
- Comparison of bispectral index scores from the standard frontal sensor position with those from an alternative mandibular positionKorean Journal of Anesthesiology, 2014
- Clinical effectiveness and cost-effectiveness of depth of anaesthesia monitoring (E-Entropy, Bispectral Index and Narcotrend): a systematic review and economic evaluationHealth Technology Assessment, 2013
- An Alternative Position for the BIS-Vista Montage in Frontal Approach Neurosurgical CasesJournal of Neurosurgical Anesthesiology, 2013
- Spectral Entropy Monitoring Reduces Anesthetic Dosage for Patients Undergoing Off-Pump Coronary Artery Bypass Graft SurgeryJournal of Cardiothoracic and Vascular Anesthesia, 2012
- BIS-Vista™ Occipital Montage in Patients Undergoing Neurosurgical Procedures during Propofol–Remifentanil AnesthesiaAnesthesiology, 2010
- Awareness in cardiac anesthesiaCurrent Opinion in Anaesthesiology, 2010
- A Comparison of Frontal and Occipital Bispectral Index Values Obtained During Neurosurgical ProceduresAnesthesia & Analgesia, 2004
- Description of the Entropy™ algorithm as applied in the Datex‐Ohmeda S/5™ Entropy ModuleActa Anaesthesiologica Scandinavica, 2004
- Bispectral index: comparison of two montagesBritish Journal of Anaesthesia, 1998
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986