Pulsed Magnetic Field Generator for Enhancing Cell Membrane Permeability In Vitro

Authors:

Abstract
The noncontact pulsed magnetic field (PMF) treatment can enhance cell membrane permeability, but its efficiency is relatively low. To enhance its efficiency, this article has developed a more optimized PMF generator: this generator can produce a maximum magnetic induction intensity of 1.9 T; and a maximum-induced electric field of approximately 110 V/cm with a uniformity of up to 84%. Importantly, the generator developed in this article can also generate a higher spatial gradient of the magnetic field, approximately 1000 T/m, a capability that was previously lacking in research. Subsequently, by customizing the coil structure and the heat dissipation system, we ensured both temperature control of the treatment plane and a sufficiently large effective treatment volume. Finally, the effectiveness of the PMF generator was confirmed through in vitro cell experiments.
Funding Information
  • National Natural Science Foundation of China (52077022)
  • 111 Project (BP0820005)

This publication has 34 references indexed in Scilit: