Energy Calibration of the Pixels of Spectral X-ray Detectors
- 14 July 2014
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 34 (3) , 697-706
Abstract
The energy information acquired using spectral X-ray detectors allows noninvasive identification and characterization of chemical components of a material. To achieve this, it is important that the energy response of the detector is calibrated. The established techniques for energy calibration are not practical for routine use in pre-clinical or clinical research environment. This is due to the requirements of using monochromatic radiation sources such as synchrotron, radio-isotopes, and prohibitively long time needed to set up the equipment and make measurements. To address these limitations, we have developed an automated technique for calibrating the energy response of the pixels in a spectral X-ray detector that runs with minimal user intervention. This technique uses the X-ray tube voltage (kVp) as a reference energy, which is stepped through an energy range of interest. This technique locates the energy threshold where a pixel transitions from not-counting (off) to counting (on). Similarly, we have developed a technique for calibrating the energy response of individual pixels using X-ray fluorescence generated by metallic targets directly irradiated with polychromatic X-rays, and additionally γ-rays from 241 Am. This technique was used to measure the energy response of individual pixels in CdTe-Medipix3RX by characterizing noise performance, threshold dispersion, gain variation and spectral resolution. The comparison of these two techniques shows the energy difference of 1 keV at 59.5 keV which is less than the spectral resolution of the detector (full-width at half-maximum of 8 keV at 59.5 keV). Both techniques can be used as quality control tools in a pre-clinical multi-energy CT scanner using spectral X-ray detectors.Keywords
This publication has 24 references indexed in Scilit:
- Imaging properties of small-pixel spectroscopic x-ray detectors based on cadmium telluride sensorsPhysics in Medicine & Biology, 2012
- Characterization of the Medipix3 pixel readout chipJournal of Instrumentation, 2011
- Characterization of Medipix3 with the MARS readout and softwareJournal of Instrumentation, 2011
- The Influence of Pixel Pitch and Electrode Pad Size on the Spectroscopic Performance of a Photon Counting Pixel Detector With CdTe SensorIEEE Transactions on Nuclear Science, 2010
- Characterization of Medipix3 With Synchrotron RadiationIEEE Transactions on Nuclear Science, 2010
- Progress in the Development of CdTe and CdZnTe Semiconductor Radiation Detectors for Astrophysical and Medical ApplicationsSensors, 2009
- Energy calibration measurements of MediPix2Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2008
- Calculation of x‐ray spectra emerging from an x‐ray tube. Part I. Electron penetration characteristics in x‐ray targetsMedical Physics, 2007
- Spectroscopy applications for the Medipix photon counting X-ray systemNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2004
- Evaluation of a photon-counting hybrid pixel detector array with a synchrotron X-ray sourceNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2002