Impact of limited temporal resolution on 4D Monte Carlo dose calculation for intensity modulated proton therapy
Authors:
Preprint
Abstract
The interplay between the beam delivery time structure and the patient motion makes 4D dose calculation (4DDC) important when treating moving tumors with intensity modulated proton therapy. 4DDC based on phase sorting of a 4DCT suffers from approximation errors in the assignment of spots to phases, since the temporal image resolution of the 4DCT is much lower than that of the delivery time structure. In this study we investigate and address this limitation by a method which applies registration-based interpolation between phase images to increase the temporal resolution of the 4DCT. First, each phase image is deformed toward its neighbor using the deformation vector field that aligns them, scaled by the desired time step. Then Monte Carlo-based 4DDC is performed on both the original 4DCT (10 phases), and extended 4DCTs at increasingly fine temporal resolutions. The method was evaluated on seven lung cancer patients treated with three robustly optimized beams, with simulated delivery time structures. Errors resulting from limited temporal resolution were measured by comparisons of doses computed using extended 4DCTs of various resolutions. The dose differences were quantified by gamma pass rates and volumes of the CTV that had dose differences above certain thresholds. The ground truth was taken as the dose computed using 100 phase images, and was justified by considering the diminishing effects of adding more images. The effect on dose-averaged linear energy transfer was also included in the analysis. A resolution of 20 (30) phase images per breathing cycle was sufficient to bring mean CTV $γ$-pass rates for $3\%/3$mm ($2\%/2$mm) above 99$\%$. For the patients with well behaved image data, mean CTV $γ$-pass rates for $1\%/1$mm surpassed 99$\%$ at a resolution of 50 images.
Keywords
All Related Versions
This publication has 21 references indexed in Scilit:
- Clinical validation of a GPU-based Monte Carlo dose engine of a commercial treatment planning system for pencil beam scanning proton therapyPhysica Medica, 2021
- 4D robust optimization including uncertainties in time structures can reduce the interplay effect in proton pencil beam scanning radiation therapyMedical Physics, 2018
- Pencil Beam Algorithms Are Unsuitable for Proton Dose Calculations in LungInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2017
- Use of image registration and fusion algorithms and techniques in radiotherapy: Report of the AAPM Radiation Therapy Committee Task Group No. 132Medical Physics, 2017
- A longitudinal four‐dimensional computed tomography and cone beam computed tomography dataset for image‐guided radiation therapy research in lung cancerMedical Physics, 2016
- The Cancer Imaging Archive (TCIA): Maintaining and Operating a Public Information RepositoryJournal of Digital Imaging, 2013
- Motion Interplay as a Function of Patient Parameters and Spot Size in Spot Scanning Proton Therapy for Lung CancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2013
- Quantification of interplay effects of scanned particle beams and moving targetsPhysics in Medicine & Biology, 2008
- Image interpolation in 4D CT using a BSpline deformable registration modelInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- Intrafractional motion during proton beam scanningPhysics in Medicine & Biology, 2005