Comparison of Enterprise Point‐of‐Care and Nova Biomedical Critical Care Xpress analyzers for determination of arterialpH, blood gas, and electrolyte values in canine and equine blood
- 9 July 2018
- journal article
- research article
- Published by Wiley in Veterinary Clinical Pathology
- Vol. 47 (3) , 415-424
Abstract
Background: Point‐of‐care analyzers can provide a rapid turnaround time for critical blood test results. Agreement between the Enterprise Point‐of‐Care (EPOC) and bench‐top laboratory analyzers is important to determine the clinical reliability of theEPOC.Objectives: The aim of the study was (1) to evaluate the precision (repeatability) of blood gas values measured by theEPOCand (2) to determine the level of agreement between theEPOCand Nova Critical Care Express (NovaCCX) for the assessment of arterialpH, blood gases, and electrolyte variables in canine and equine blood.Methods: Arterial blood samples from dogs were analyzed on theEPOCand NovaCCXanalyzers to determine precision and agreement ofpH, PaCO2, PaO2, andHCT. The same analytes plus Na+, K−, and Cl−were analyzed for agreement using equine blood. Statistical analyses included assessment of precision using the coefficient of variation (CV%), and agreement using the Deming regression, Pearson correlation, and Bland‐Altman plots.Results: Both analyzers provided precise results ofpH, PaCO2, PaO2,andHCT, meetingCV% quality requirement values. In both species, Deming regression results were acceptable and correlation values were above 0.93 for arterialpHand blood gases, but lower for sodium and chloride. Bland‐Altman plots demonstrated varying degrees of bias, but good agreement between the 2 analyzers was seen when arterial blood gases and electrolytes were measured, except for PaCO2and Cl−.Conclusion: TheEPOCanalyzer provides consistent, reliable results for canine arterial blood gas values and for equine arterial blood gas and electrolyte values. Cl−results could be acceptable with the application of a correction factor, but the PaCO2results were more variable.Keywords
Funding Information
- Washington State University
This publication has 31 references indexed in Scilit:
- Clinical Evaluation of the Enterprise Point of Care for Blood Gas Electrolytes and MetabolitesPoint of Care: The Journal of Near-Patient Testing & Technology, 2013
- Ion Selective Electrodes (ISEs) and interferences—A reviewClinica Chimica Acta, 2010
- Design, Analysis, and Interpretation of Method-Comparison StudiesAACN Advanced Critical Care, 2008
- Evaluation of the Enterprise Point-of-Care (EPOC) System for Blood Gas and Electrolyte AnalysisPoint of Care: The Journal of Near-Patient Testing & Technology, 2008
- Use of the i-STAT portable clinical analyzer in miceLab Animal, 2006
- Heparin interference in whole blood sodium measurements in a pediatric settingClinical Biochemistry, 2006
- Evaluation of the i‐STAT hand‐held chemical analyser during treadmill and endurance exerciseEquine Veterinary Journal, 2002
- Reliability: What is it, and how is it measured?Physiotherapy, 2000
- Current databases on biological variation: pros, cons and progressScandinavian Journal of Clinical and Laboratory Investigation, 1999
- A clinical view of analytical goals in clinical biochemistry.Journal of Clinical Pathology, 1979