Evaluating NIR vascular imaging to support intravenous cannulation in awake children difficult to cannulate; a randomized clinical trial
- 2 August 2014
- journal article
- research article
- Published by Wiley in Pediatric Anesthesia
- Vol. 24 (11) , 1174-1179
Abstract
Summary: Background: Recently, various near‐infrared vascular imaging devices aimed at facilitating peripheral intravenous cannulation (PIC) were introduced, all claiming to increase success rate of PIC. We evaluated the clinical utility of a near‐infrared vascular imaging device (VascuLuminator®) in pediatric patients who were referred to the anesthesiologist because of difficult cannulation.Methods: There were 226 consecutive children referred to pediatric anesthesiologists by the treating pediatrician of the in‐ and outpatient clinic, because of difficulties with intravenous cannulation, were included in this cluster randomized clinical trial. The presence and use of the near‐infrared vascular imaging device for PIC was randomized in clusters of 1 week. Success at first attempt (Fisher exact test) and time to successful cannulation (Log‐rank test) were assessed to evaluate differences between groups.Results: Success at first attempt in the group with the VascuLuminator® (59%) was not significantly different from the control group (54%, P = 0.41), neither was the median time to successful cannulation: 246 s and 300 s, respectively (P = 0.54).Conclusions: Visualization of blood vessels with near‐infrared light and with near‐infrared vascular imaging device did not improve success of PIC in pediatric patients who are known difficult to cannulate.Keywords
Funding Information
- ZonMw Netherlands Organization for Health Research and Development (170991008)
This publication has 20 references indexed in Scilit:
- Near‐infrared light to aid peripheral intravenous cannulation in children: a cluster randomised clinical trial of three devicesAnaesthesia, 2013
- Near-Infrared Imaging in Intravenous Cannulation in Children: A Cluster Randomized Clinical TrialPediatrics, 2013
- Does the Use of an Assistive Device by Nurses Impact Peripheral Intravenous Catheter Insertion Success in Children?Journal of Pediatric Nursing, 2012
- Efficacy of VeinViewer in pediatric peripheral intravenous access: a randomized controlled trialEuropean Journal of Pediatrics, 2012
- A randomized trial of the Vein Viewer versus standard technique for placement of peripherally inserted central catheters (PICCs) in neonatesJournal of Perinatology, 2011
- VeinViewer‐assisted Intravenous Catheter Placement in a Pediatric Emergency DepartmentAcademic Emergency Medicine, 2011
- Predictive factors for difficult intravenous cannulation in pediatric patients at a tertiary pediatric hospitalPediatric Anesthesia, 2011
- Efficacy of a Near-Infrared Light Device in Pediatric Intravenous CannulationPediatric Emergency Care, 2011
- Clinical Implications of Unmanaged Needle-Insertion Pain and Distress in ChildrenPediatrics, 2008
- Veinlite Transillumination in the Pediatric Emergency DepartmentPediatric Emergency Care, 2008