Ghosting caused by bulk charge trapping in direct conversion flat-panel detectors using amorphous selenium
Authors:
- Wei ZhaoWZWei Zhao
- G. DeCrescenzoGDG. DeCrescenzo
- Safa O. KasapSKSafa O. Kasap
- J. A. RowlandsJRJ. A. Rowlands
- 27 January 2005
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 32 (2) , 488-500
Abstract
Direct flat-panel detectors using amorphous selenium (a-Se) x-ray photoconductors are gaining wide-spread clinical use. The goal of our investigation is to understand the physical mechanisms responsible for ghosting, i.e., x-ray induced change in sensitivity that results in image persistence, so that the knowledge can be used to consistently minimize ghosting artifacts in a-Se flat-panel detectors. In this paper we will discuss the effect on x-ray sensitivity of charge trapping in a-Se, which is the dominant source for ghosting in a-Se flat-panel detectors. Our approach is to correlate ghosting in electroded a-Se detectors with the trapped charge concentration measured by the "time-of-flight" (TOF) method. All measurements were performed as a function of radiation exposure X of up to approximately 20 R at electric field strength's of E(Se)=5 and 10 V/microm. The results showed that the x-ray sensitivity decreased as a function of X and the amount of ghosting decreased with increasing E(Se). The shape of the TOF curves changed as a result of irradiation in a manner indicating trapped electrons in the bulk of a-Se. The density of trapped electrons n(t) increases as a function of X. A method was developed to determine the values of n(t) in the bulk of a-Se from the TOF measurements, and to predict the corresponding change in x-ray sensitivity. Our results showed that a recombination coefficient consistent with that predicted by Langevin produced good agreement between calculated and measured x-ray sensitivity changes. Thus it can be concluded that the trapping of electrons in the bulk of a-Se and their subsequent recombination with x-ray generated free holes is the dominant mechanism for ghosting in a-Se.Keywords
This publication has 39 references indexed in Scilit:
- Sensitivity reduction in biased amorphous selenium photoconductorsJournal of Physics D: Applied Physics, 2002
- X-ray induced effects in stabilized a-Se X-ray photoconductorsJournal of Non-Crystalline Solids, 2002
- X-ray irradiation induced bulk space charge in stabilized a-Se x-ray photoconductorsJournal of Applied Physics, 1998
- Non-Langevin recombination in disordered materials with random potential distributionsSolid State Communications, 1997
- X‐ray imaging using amorphous selenium: Photoinduced discharge (PID) readout for digital general radiographyMedical Physics, 1995
- X-ray imaging using amorphous selenium: Determination of x-ray sensitivity by pulse height spectroscopyMedical Physics, 1992
- A new photoconductor imaging system for digital radiographyMedical Physics, 1989
- Transient photoconductivity probing of negative bulk space charge evolution in halogenated amorphous selenium filmsSolid State Communications, 1987
- The mechanism of sensitivity reduction in selenium layers irradiated by X-raysPhysica Status Solidi (a), 1974
- XeroradiographyPhysics in Medicine & Biology, 1973