Precision and safety of Multilevel Cervical Transpedicular Screw Fixation with 3D Patient-Specific Guides; A Cadaveric Study
Open Access
- 30 October 2019
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 9 (1) , 1-7
Abstract
The aim is to design a patient-specific instrument (PSI) for multilevel cervical pedicle screw placement from C2 to C7, as well as verifying reliability and reproducibility. Computed tomography (CT) scans were obtained from 7 cadaveric cervical spines. Using Mimics software, semiautomatic segmentation was performed for each cervical spine, designing a 3D cervical spine bone model in order to plan transpedicular screw fixation. A PSI was designed according to the previously cited with two cannulated chimneys to guide the drill. The guides were 3D printed and surgeries performed at the laboratory. Postoperative scans were obtained to study screw placement. Sixty-eight transpedicular screws were available for study. 61.8% of all screws were within the pedicle or partially breached <4 mm. No differences were observed between cervical levels. None of these screws had neurovascular injury. Of the 27 screws with a grade 3 (screw outside the pedicle; 39.7%), only 2 had perforation of the transverse foramen and none of them would have caused a neural injury. In conclusion, multilevel PSI for cervical pedicle screw is a promising technology that despite showing improvements regarding free-hand technique requires further studies to improve the positioning of the PSI and their accuracy.Keywords
This publication has 28 references indexed in Scilit:
- Intraoperative, full-rotation, three-dimensional image (O-arm)–based navigation system for cervical pedicle screw insertionJournal of Neurosurgery: Spine, 2011
- Accuracy of pedicle screw placement: a systematic review of prospective in vivo studies comparing free hand, fluoroscopy guidance and navigation techniquesEuropean Spine Journal, 2011
- Vertebral artery injury in cervical spine surgery: anatomical considerations, management, and preventive measuresThe Spine Journal, 2009
- Comparison of two novel fluoroscopy-based stereotactic methods for cervical pedicle screw placement and review of the literatureEuropean Spine Journal, 2008
- Lateral mass screw fixation for cervical spine trauma: associated complications and efficacy in maintaining alignmentThe Spine Journal, 2006
- Cervical Pedicle Screws: Conventional Versus Computer-Assisted Placement of Cannulated ScrewsSpine, 2005
- Pedicle Screws Enhance Primary Stability in Multilevel Cervical Corpectomies: Biomechanical In Vitro Comparison of Different Implants Including Constrained and Nonconstrained Posterior InstumentationsSpine, 2003
- Atlantoaxial Fixation Using Plate and Screw Method: A Report of 160 Treated PatientsNeurosurgery, 2002
- Placement of Pedicle Screws in the Human Cadaveric Cervical SpineSpine, 2000
- Transpedicular Screw Fixation for Traumatic Lesions of the Middle and Lower Cervical SpineJournal of Spinal Disorders, 1994