How to assess intra- and inter-observer agreement with quantitative PET using variance component analysis: a proposal for standardisation
Open Access
- 21 September 2016
- journal article
- research article
- Published by Springer Nature in BMC Medical Imaging
- Vol. 16 (1) , 1-9
Abstract
Quantitative measurement procedures need to be accurate and precise to justify their clinical use. Precision reflects deviation of groups of measurement from another, often expressed as proportions of agreement, standard errors of measurement, coefficients of variation, or the Bland-Altman plot. We suggest variance component analysis (VCA) to estimate the influence of errors due to single elements of a PET scan (scanner, time point, observer, etc.) to express the composite uncertainty of repeated measurements and obtain relevant repeatability coefficients (RCs) which have a unique relation to Bland-Altman plots. Here, we present this approach for assessment of intra- and inter-observer variation with PET/CT exemplified with data from two clinical studies. In study 1, 30 patients were scanned pre-operatively for the assessment of ovarian cancer, and their scans were assessed twice by the same observer to study intra-observer agreement. In study 2, 14 patients with glioma were scanned up to five times. Resulting 49 scans were assessed by three observers to examine inter-observer agreement. Outcome variables were SUVmax in study 1 and cerebral total hemispheric glycolysis (THG) in study 2. In study 1, we found a RC of 2.46 equalling half the width of the Bland-Altman limits of agreement. In study 2, the RC for identical conditions (same scanner, patient, time point, and observer) was 2392; allowing for different scanners increased the RC to 2543. Inter-observer differences were negligible compared to differences owing to other factors; between observer 1 and 2: −10 (95 % CI: −352 to 332) and between observer 1 vs 3: 28 (95 % CI: −313 to 370). VCA is an appealing approach for weighing different sources of variation against each other, summarised as RCs. The involved linear mixed effects models require carefully considered sample sizes to account for the challenge of sufficiently accurately estimating variance components.Keywords
This publication has 41 references indexed in Scilit:
- Evaluation of CT‐based lean‐body SUVMedical Physics, 2013
- Repeatable Noninvasive Measurement of Mouse Myocardial Glucose Uptake with 18F-FDG: Evaluation of Tracer Kinetics in a Type 1 Diabetes ModelJournal of Nuclear Medicine, 2013
- Statistical Methods Used to Test for Agreement of Medical Instruments Measuring Continuous Variables in Method Comparison Studies: A Systematic ReviewPLOS ONE, 2012
- Guidelines for Reporting Reliability and Agreement Studies (GRRAS) were proposedJournal of Clinical Epidemiology, 2011
- FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0European Journal of Nuclear Medicine and Molecular Imaging, 2009
- Reproducibility of Tumor Blood Flow Quantification with 15O-Water PETJournal of Nuclear Medicine, 2008
- When to use agreement versus reliability measuresJournal of Clinical Epidemiology, 2006
- Measuring change over time: Assessing the usefulness of evaluative instrumentsJournal of Chronic Diseases, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Regression modelling strategies for improved prognostic predictionStatistics in Medicine, 1984