Motorization of a surgical microscope for intra‐operative navigation and intuitive control
Authors:
- 26 August 2010
- journal article
- research article
- Published by Wiley in The International Journal of Medical Robotics and Computer Assisted Surgery
- Vol. 6 (3) , 269-280
Abstract
Background: During surgical procedures, various medical systems, e.g. microscope or C‐arm, are used. Their precise and repeatable manual positioning can be very cumbersome and interrupts the surgeon's work flow. Robotized systems can assist the surgeon but they require suitable kinematics and control. However, positioning must be fast, flexible and intuitive.Methods: We describe a fully motorized surgical microscope. Hardware components as well as implemented applications are specified. The kinematic equations are described and a novel control concept is proposed.Results: Our microscope combines fast manual handling with accurate, automatic positioning. Intuitive control is provided by a small remote control mounted to one of the surgical instruments. Positioning accuracy and repeatability are < 1 mm and vibrations caused by automatic movements fade away in about 1 s.Conclusion: The robotic system assists the surgeon, so that he can position the microscope precisely and repeatedly without interrupting the clinical workflow. The combination of manual und automatic control guarantees fast and flexible positioning during surgical procedures. Copyright © 2010 John Wiley & Sons, Ltd.Keywords
This publication has 13 references indexed in Scilit:
- Combining Optical Coherence Tomography (OCT) with an Operating Microscope2007
- A robotic C-arm fluoroscopeThe International Journal of Medical Robotics and Computer Assisted Surgery, 2005
- Pivoting motion control for a laparoscopic assistant robot and human clinical trialsAdvanced Robotics, 2005
- ENTWICKLUNG EINER KOMPAKTEN, TEILROBOTISCHEN TRÄGERPLATTFORM FÜR EIN ELEKTRONISCHES OP-MIKROSKOPBiomedical Engineering / Biomedizinische Technik, 2002
- Mouth Operated Neurosurgical Robot2001
- Real-Time Inverse Kinematics Techniques for Anthropomorphic LimbsGraphical Models, 2000
- Experience With Various 3-Dimensional Navigation Systems in Head and Neck SurgeryJAMA Otolaryngology–Head & Neck Surgery, 2000
- Modelling and Control of Robot Manipulators2000
- Optical Coherence TomographyScience, 1991
- Closed-form solution of absolute orientation using unit quaternionsJournal of the Optical Society of America A, 1987