Global Thrombosis Test: Occlusion by Coagulation or SIPA?
Open Access
- 19 September 2021
- journal article
- research article
- Published by Georg Thieme Verlag KG in TH Open
- Vol. 05 (03) , e400-e410
Abstract
The global thrombosis test (GTT) is a point of care device that tests thrombotic and thrombolytic status. The device exposes whole blood flow to a combination of both high and low shear stress past and between ball bearings potentially causing thrombin and fibrin formation. The question arises as to whether thrombosis in the GTT is dominated by coagulation-triggered red clot or high shear-induced white clot. We investigated the nature of the thrombus formed in the GTT, the device efficacy, human factors use, and limitations. The GTT formed clots that were histologically fibrin-rich with trapped red blood cells. The occlusion time (OT) was more consistent with coagulation than high shear white clot and was strongly lengthened by heparin and citrate, two common anticoagulants. The clot was lysed by tissue plasminogen activator (tPA), also consistent with a fibrin-rich red clot. Changing the bead to a collagen-coated surface and eliminating the low shear zone between the beads induced a rapid OT consistent with a platelet-rich thrombus that was relatively resistant to heparin or tPA. The evidence points to the GTT as occluding primarily due to fibrin-rich red clot from coagulation rather than high shear platelet aggregation and occlusion associated with arterial thrombosis. * V.C. and P.J.Y. contributed equally as co-first authors. Received: 18 February 2021 Accepted: 16 June 2021 Article published online: 19 September 2021 © 2021. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/) Georg Thieme Verlag KG Rüdigerstraße 14, 70469 Stuttgart, GermanyKeywords
Funding Information
- Atlanta Center for Microsystems Engineered Point-of-Care Technologies, ACME-POCT (NIH-5U54EB027690)
This publication has 37 references indexed in Scilit:
- Sources of Variability in Platelet Accumulation on Type 1 Fibrillar Collagen in Microfluidic Flow AssaysPLOS ONE, 2013
- Impaired spontaneous thrombolytic activity measured by global thrombosis test in males with metabolic syndromeThrombosis Research, 2012
- Rapid Platelet Accumulation Leading to Thrombotic OcclusionAnnals of Biomedical Engineering, 2011
- Platelet Function and Spontaneous Thrombolytic Activity of Patients with Cerebral Infarction Assessed by the Global Thrombosis TestPathophysiology of Haemostasis and Thrombosis, 2010
- The growing complexity of platelet aggregationBlood, 2007
- Gorog Thrombosis Test: analysis of factors influencing occlusive thrombus formationBlood Coagulation & Fibrinolysis, 2006
- Impaired spontaneous thrombolytic activity in elderly and in habitual smokers, as measured by a new global thrombosis testBlood Coagulation & Fibrinolysis, 2003
- Görög Thrombosis TestBlood Coagulation & Fibrinolysis, 2003
- Orcein-Hematoxylin in Iodized Ferric Chloride as a Stain for Elastic Fibers, With Metanil Yellow CounterstainingStain Technology, 1967
- The identification of platelets and platelet antigens in histological sectionsThe Journal of Pathology and Bacteriology, 1965