A wireless accelerometer node for reliable and valid measurement of lumbar accelerations during treadmill running
- 2 January 2016
- journal article
- research article
- Published by Taylor & Francis in Sports Biomechanics
- Vol. 15 (1) , 11-22
Abstract
This study investigated the reliability of a wireless accelerometer and its agreement with optical motion capture for the measurement of root mean square (RMS) acceleration during running. RMS acceleration provides a whole-body metric of movement mechanics and economy. Fifteen healthy college-age participants performed treadmill running for two 60-s trials at 2.22, 2.78, and 3.33 m/s and one trial of 150 s (five 30-s epochs) at 2.78 m/s. We assessed between-trial and within-trial reliability, and agreement in each axis between a trunk-mounted wireless accelerometer and a reflective marker on the accelerometer measured by optical motion capture. Intraclass correlations assessing between-trial repeatability were 0.89–0.97, depending on the axis, and intraclass correlations assessing within-trial repeatability were 0.99–1.00. Bland–Altman analyses assessing agreement indicated mean difference values between −0.03 and 0.03 g, depending on the axis. Anterio-posterior acceleration had the greatest limits of agreement (LOA) (±0.12 g) and vertical acceleration had the smallest LOA (±0.03 g). For measuring RMS acceleration of the trunk, this wireless accelerometer node provides repeatable and valid measurement compared with the standard laboratory method of optical motion capture.Keywords
This publication has 21 references indexed in Scilit:
- Ground Contact Time as an Indicator of Metabolic Cost in Elite Distance RunnersMedicine & Science in Sports & Exercise, 2012
- Calibration and Validation of Wearable MonitorsMedicine & Science in Sports & Exercise, 2012
- Assessing Physical Activity Using Wearable MonitorsMedicine & Science in Sports & Exercise, 2012
- High Resolution MEMS Accelerometers to Estimate VO2 and Compare Running Mechanics between Highly Trained Inter-Collegiate and Untrained RunnersPLOS ONE, 2009
- Control entropy identifies differential changes in complexity of walking and running gait patterns with increasing speed in highly trained runnersChaos: An Interdisciplinary Journal of Nonlinear Science, 2009
- Walking and MeasurementMedicine & Science in Sports & Exercise, 2008
- Acceleration versus heart rate for estimating energy expenditure and speed during locomotion in animals: Tests with an easy model species, Homo sapiensZoology, 2008
- Estimation of Oxygen Uptake during Fast Running Using Accelerometry and Heart RateMedicine & Science in Sports & Exercise, 2007
- Test–retest reliability of trunk accelerometric gait analysisGait & Posture, 2004
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986