A Method of Detecting Heartbeat Locations in the Ballistocardiographic Signal From the Fiber-Optic Vital Signs Sensor
Authors:
Open Access
- 23 January 2015
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Journal of Biomedical and Health Informatics
- Vol. 19 (4) , 1443-1450
Abstract
We present a flexible, easy-to-expand digital signal processing method for detecting heart rate (HR) for cardiac vibration signals of fiber Bragg grating (FBG) sensor. The FBG-based method of measuring HR is possible to use during the magnetic resonance imaging procedure, which is its unique advantage. Our goal was to design a detection method with plurality of parameters and to subject these parameters to genetic algorithm optimization technique. In effect, we arrived at a method that is well able to deal with much distorted signals with low SNR. We proved that the method we developed allows automatic adjustment to the shape of the waves of signal carrying useful information about the moments of heartbeat. Thus, we can easily adapt our technique to the analysis of signals, which contains information on HR, from sensors employing different techniques of strain detection. The proposed method has the capabilities of analyzing signals in semi-real-time (online) with beat-to-beat resolution, significantly low delay, and negligible computational power requirements. We verified our method on recordings in a group of seven subjects. Verification included over 6000 heartbeats (82 min 47 s of recordings). The root-mean-square error of our method does not exceed 6.0 bpm.Keywords
Funding Information
- National Defense and Security (DOBR/0052/R/ID1/2012/03)
This publication has 23 references indexed in Scilit:
- Simultaneous measurement of breathing rate and heart rate using a microbend multimode fiber optic sensorJournal of Biomedical Optics, 2014
- A new approach to ballistocardiographic measurements using fibre Bragg grating-based sensorsBiocybernetics and Biomedical Engineering, 2014
- Optimization of Heartbeat Detection in Fiber-Optic Unobtrusive Measurements by Using Maximum A Posteriori Probability EstimationIEEE Journal of Biomedical and Health Informatics, 2013
- Fiber Bragg grating-based sensor for monitoring respiration and heart activity during magnetic resonance imaging examinationsJournal of Biomedical Optics, 2013
- Robust inter-beat interval estimation in cardiac vibration signalsPhysiological Measurement, 2013
- Portable fiber optic ballistocardiogram sensor for home use2012
- Medical Textiles With Embedded Fiber Optic Sensors for Monitoring of Respiratory MovementIEEE Sensors Journal, 2011
- Adaptive Beat-to-Beat Heart Rate Estimation in BallistocardiogramsIEEE Transactions on Information Technology in Biomedicine, 2011
- Optical Fiber Sensors Embedded Into Medical Textiles for Healthcare MonitoringIEEE Sensors Journal, 2008
- Handbook of Evolutionary Computation1997