Consistency conditions for cone-beam CT data acquired with a straight-line source trajectory
Authors:
- Margo S LevineMLMargo S Levine
- Emil Y SidkyESEmil Y Sidky
- Xiaochuan PanXPXiaochuan Pan
Open Access
- 1 February 2010
- journal article
- Published by Tsinghua University Press in Tsinghua Science and Technology
- Vol. 15 (1) , 56-61
Abstract
A consistency condition is developed for computed tomography (CT) projection data acquired from a straight-line X-ray source trajectory. The condition states that integrals of normalized projection data along detector lines parallel to the X-ray path must be equal. The projection data is required to be untruncated only along the detector lines parallel to the X-ray path, a less restrictive requirement compared to Fourier conditions that necessitate completely untruncated data. The condition is implemented numerically on simple image functions, a discretization error bound is estimated, and detection of motion inconsistencies is demonstrated. The results show that the consistency condition may be used to quantitatively compare the quality of projection data sets obtained from different scans of the same image object.Keywords
This publication has 5 references indexed in Scilit:
- Consistency conditions upon 3D CT data and the wave equationPhysics in Medicine & Biology, 2002
- A fast rebinning algorithm for 3D positron emission tomography using John's equationInverse Problems, 1999
- Mathematical framework of cone beam 3D reconstruction via the first derivative of the radon transform1991
- Cone Beam Reconstruction with Sources on a CurveSIAM Journal on Applied Mathematics, 1985
- The ultrahyperbolic differential equation with four independent variablesDuke Mathematical Journal, 1938