Intrarenal manipulation of flexible ureteroscopes: a comparative study

Authors:

Abstract
OBJECTIVE: To compare the relative ease of manipulation around a calyceal model in the hands of experienced ureteroscopists of four flexible ureteroscopes (FUs), as new‐generation FUs provide exaggerated active deflection to facilitate intrarenal manipulation. SUBJECTS AND METHODS: Four FUs of P = 0.034). The URF‐P3 took longer to reach the posterior lower pole, at 18.1 (14.5) s than the Viper, SF and D8E, at 6.3 (2.8) ( P = 0.004), 6.6 (3.3) ( P = 0.008) and 7.7 (4.9) ( P = 0.008), respectively. The Viper was quicker to reach the posterior middle pole, at 7.3 (4.0) s, than the URF‐P3, SF and D8E, at 14.9 (13.1) ( P = 0.036), 14.7 (14.3) ( P = 0.056) and 12.7 (14.0) s ( P = 0.115), respectively. Subjectively, six of eight endoscopists stated that the Viper was the better FU for ease of manipulation, while five of eight stated that the URF‐P3 was the most difficult. Notably, all eight endoscopists had more clinical experience with the URF‐P3 than with the other three FUs tested. CONCLUSIONS: In a model of a moderately hydronephrotic kidney with long infundibulum and limited ability to perform passive secondary deflection, FUs with exaggerated active deflection outmanoeuvred the URF‐P3. The Viper was better for manipulation in the hands of experienced endoscopists.