Validation of the quantitative point-of-care CareStart biosensor for assessment of G6PD activity in venous blood
Open Access
- 8 May 2018
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 13 (5) , e0196716
Abstract
Glucose-6-phosphate dehydrogenase (G6PD) deficiency is the most common enzymopathy in the human population affecting an estimated 8% of the world population, especially those living in areas of past and present malaria endemicity. Decreased G6PD enzymatic activity is associated with drug-induced hemolysis and increased risk of severe neonatal hyperbilirubinemia leading to brain damage. The G6PD gene is on the X chromosome therefore mutations cause enzymatic deficiency in hemizygote males and homozygote females while the majority of heterozygous females have an intermediate activity (between 30–80% of normal) with a large distribution into the range of deficiency and normality. Current G6PD qualitative tests are unable to diagnose G6PD intermediate activities which could hinder wide use of 8-aminoquinolines for Plasmodium vivax elimination. The aim of the study was to assess the diagnostic performances of the new Carestart G6PD quantitative biosensor. A total of 150 samples of venous blood with G6PD deficient, intermediate and normal phenotypes were collected among healthy volunteers living along the north-western Thailand-Myanmar border. Samples were analyzed by complete blood count, by gold standard spectrophotometric assay using Trinity kits and by the latest model of Carestart G6PD biosensor which analyzes both G6PD and hemoglobin. Bland-Altman comparison of the CareStart normalized G6PD values to that of the gold standard assay showed a strong bias in values resulting in poor area under-the-curve values for both 30% and 80% thresholds. Performing a receiver operator curve identified threshold values for the CareStart product equivalent to the 30% and 80% gold standard values with good sensitivity and specificity values, 100% and 92% (for 30% G6PD activity) and 92% and 94% (for 80% activity) respectively. The Carestart G6PD biosensor represents a significant improvement for quantitative diagnosis of G6PD deficiency over previous versions. Further improvements and validation studies are required to assess its utility for informing radical cure decisions in malaria endemic settings.Keywords
Funding Information
- Bill and Melinda Gates Foundation (OPP1034534)
- Department for International Development (204139)
This publication has 32 references indexed in Scilit:
- G6PD testing in support of treatment and elimination of malaria: recommendations for evaluation of G6PD testsMalaria Journal, 2013
- Neonatal Intensive Care in a Karen Refugee Camp: A 4 Year Descriptive StudyPLOS ONE, 2013
- Randomized, Open-Label Trial of Primaquine against Vivax Malaria Relapse in IndonesiaAntimicrobial Agents and Chemotherapy, 2013
- Clinical spectrum and severity of hemolytic anemia in glucose 6-phosphate dehydrogenase–deficient children receiving dapsoneBlood, 2012
- G6PD Deficiency Prevalence and Estimates of Affected Populations in Malaria Endemic Countries: A Geostatistical Model-Based MapPLOS Medicine, 2012
- Relapses Contribute Significantly to the Risk of Plasmodium vivax Infection and Disease in Papua New Guinean Children 1–5 Years of AgeThe Journal of Infectious Diseases, 2012
- Medications and Glucose-6-Phosphate Dehydrogenase DeficiencyDrug Safety, 2010
- Glucose-6-phosphate dehydrogenase deficiencyThe Lancet, 2008
- Flow-Cytometric Analysis of Reticulocytes in Normal Cord BloodActa Haematologica, 1993
- International Committee for Standardization in Haematology: Recommended Methods for Red‐Cell Enzyme Analysis*British Journal of Haematology, 1977