A Clue to the Existence of Bonding between Bone and Implant Surface: An In Vivo Study
Open Access
Abstract
We evaluated the shear bond strength of bone–implant contact, or osseointegration, in the rabbit tibia model, and compared the strength between grades 2 and 4 of commercially pure titanium (cp-Ti). A total of 13 grades 2 and 4 cp-Ti implants were used, which had an identical cylinder shape and surface topography. Field emission scanning electron microscopy, X-ray photoelectron spectroscopy, and confocal laser microscopy were used for surface analysis. Four grades 2 and 4 cp-Ti implants were inserted into the rabbit tibiae with complete randomization. After six weeks of healing, the experimental animals were sacrificed and the implants were removed en bloc with the surrounding bone. The bone–implant interfaces were three-dimensionally imaged with micro-computed tomography. Using these images, the bone–implant contact area was measured. Counterclockwise rotation force was applied to the implants for the measurement of removal torque values. Shear bond strength was calculated from the measured bone–implant contact and removal torque data. The t-tests were used to compare the outcome measures between the groups, and statistical significance was evaluated at the 0.05 level. Surface analysis showed that grades 2 and 4 cp-Ti implants have similar topographic features. We found no significant difference in the three-dimensional bone–implant contact area between these two implants. However, grade 2 cp-Ti implants had a higher shear bond strength than grade 4 cp-Ti implants (p = 0.032). The surfaces of the grade 2 cp-Ti implants were similar to those of the grade 4 implants in terms of physical characteristics and the quantitative amount of attachment to the bone, whereas the grade 2 surfaces were stronger than the grade 4 surfaces in the bone–surface interaction, indicating osseointegration quality.Keywords
Funding Information
- National Research Foundation of Korea (NRF-2016R1A2B4014330)
This publication has 13 references indexed in Scilit:
- Dental Implants: The Last 100 YearsJournal of Oral and Maxillofacial Surgery, 2018
- Does Change in Thread Shape Influence the Pull Out Strength of Mini Implants? An In vitro StudyJournal of Clinical and Diagnostic Research, 2017
- Importance of the Roughness and Residual Stresses of Dental Implants on Fatigue and Osseointegration Behavior. In Vivo Study in RabbitsJournal of Oral Implantology, 2016
- Establishment of a new pull-out strength testing method to quantify early osseointegration—An experimental pilot studyJournal of Cranio-Maxillofacial Surgery, 2015
- Potential for acceleration of bone formation after implant surgery by using a dietary supplement: an animal studyJournal of Oral Rehabilitation, 2015
- Reality of Dental Implant Surface Modification: A Short Literature ReviewThe Open Biomedical Engineering Journal, 2014
- Removal torque analysis of chemically modified hydrophilic and anodically oxidized titanium implants with constant angular velocity for early bone response in rabbit tibiaTissue Engineering and Regenerative Medicine, 2013
- Evaluation of Early Bone Response to Fluoride-Modified and Anodically Oxidized Titanium Implants Through Continuous Removal Torque AnalysisImplant Dentistry, 2012
- Effects of thread shape on the pullout strength of miniscrewsAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2012
- Improving bioscience research reporting: the ARRIVE guidelines for reporting animal researchOsteoarthritis and Cartilage, 2012