Abstract
In a combined positron emission tomography (PET) and computed tomography (CT) system, the CT images can be used for attenuation correction as well as for image fusion. However, quantitative and qualitative differences have been reported between CT based attenuation corrected PET and conventional transmission scan corrected PET images. The purpose of this study was to investigate potential differences in PET/CT caused by attenuation differences in bowel due to motion. Twelve patients had PET/CT scans performed using 68Ge transmission and CT attenuation correction methods. Three emission imaging datasets were generated including CT corrected PET, Ge corrected PET, and the difference images (CT corrected PET minus Ge corrected PET). PET difference images were used to identify regions of mismatch and to quantify possible discordance between images by using standardized uptake values (SUVs). Using the Ge corrected PET as the standard, differences in emission images were classified as an overestimation (pattern A) or an underestimation (pattern B) in these difference images. One hundred and twenty-three mismatched areas were identified. Among them, overestimated areas in CT corrected image were detected in 36 regions (pattern A), while underestimated areas were evaluated in the remaining 87 regions (pattern B). The mean value of the difference in pattern A (mean±standard deviation=0.84±0.44) was slightly higher than that in pattern B (0.60±0.23), and statistically significant. Six of 36 regions in pattern A had an SUV of greater than 2.5 in CT corrected PET but less than 2.5 in Ge corrected PET; two of 87 regions with pattern B demonstrated an SUV greater than 2.5 in Ge corrected PET and less than 2.5 in CT corrected PET. Physiological bowel motion may result in attenuation differences and subsequent differences in SUVs. Overestimation of fluorodeoxyglucose uptake should not be misinterpreted as disease.Keywords
This publication has 15 references indexed in Scilit:
- Accuracy of image fusion of normal upper abdominal organs visualized with PET/CTEuropean Journal of Nuclear Medicine and Molecular Imaging, 2003
- Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CTEuropean Journal of Nuclear Medicine and Molecular Imaging, 2003
- PET-CT image co-registration in the thorax: influence of respirationEuropean Journal of Nuclear Medicine and Molecular Imaging, 2002
- CT vs 68Ge attenuation correction in a combined PET/CT system: evaluation of the effect of lowering the CT tube currentEuropean Journal of Nuclear Medicine and Molecular Imaging, 2001
- PET-CT–guided Percutaneous Puncture of an Infected Cyst in Autosomal Dominant Polycystic Kidney Disease: Case ReportRadiology, 2001
- Clinical role of positron emission tomography in oncologyThe Lancet Oncology, 2001
- Combined PET/CT Imaging in Oncology Impact on Patient ManagementClinical Positron Imaging, 2000
- PET imaging in oncologySeminars in Nuclear Medicine, 2000
- Pitfalls in Oncologic Diagnosis with FDG PET Imaging: Physiologic and Benign VariantsRadioGraphics, 1999
- Normal physiological and benign pathological variants of 18-fluoro-2-deoxyglucose positron-emission tomography scanning: Potential for error in interpretationSeminars in Nuclear Medicine, 1996