Clinical Validation of Respiratory Rate Estimation Using Acoustic Signals from a Wearable Device
Open Access
- 27 November 2024
- journal article
- research article
- Published by MDPI AG in Journal of Clinical Medicine
- Vol. 13 (23) , 7199
Abstract
Objectives: Respiratory rate (RR) is a clinical measure of breathing frequency, a vital metric for clinical assessment. However, the recording and documentation of RR are considered to be extremely poor due to the limitations of the current approaches to measuring RR, including capnography and manual counting. We conducted a validation of the automatic RR measurement capability of AcuPebble RE100 (Acurable, London, UK) against a gold-standard capnography system and a type-III cardiorespiratory polygraphy system in two independent prospective and retrospective studies. Methods: The experiment for the prospective study was conducted at Imperial College London. Data from AcuPebble RE100 (Acurable, London, UK) and the reference capnography system (Capnostream™35, Medtronic, Minneapolis, MN, USA) were collected simultaneously from healthy volunteers. The data from a previously published study were used in the retrospective study, where the patients were recruited consecutively from a standard Obstructive Sleep Apnea (OSA) diagnostic pathway in a UK hospital. Overnight data during sleep were collected using the AcuPebble SA100 (Acurable, London, UK) sensor and a type-III cardiorespiratory polygraphy system (Embletta MPR Sleep System, Natus Medical, Pleasanton, CA, USA) at the patients’ homes. Data from 15 healthy volunteers were used in the prospective study. For the retrospective study, 150 consecutive patients had been referred for OSA diagnosis and successfully completed the study. Results: The RR output of AcuPebble RE100 (Acurable, London, UK) was compared against the reference device in terms of the Root Mean Squared Deviation (RMSD), mean error, and standard deviation (SD) of the difference between the paired measurements. In both the prospective and retrospective studies, the AcuPebble RE100 algorithms provided accurate RR measurements, well within the clinically relevant margin of error, typically used by FDA-approved respiratory rate monitoring devices, with the RMSD under three breaths per minute (BPM) and mean errors of 1.83 BPM and 1.4 BPM, respectively. Conclusions: The evaluation results provide evidence that AcuPebble RE100 (Acurable, London, UK) algorithms produce reliable results and are hence suitable for overnight monitoring of RR.Keywords
Funding Information
- European Research Council (724334)
This publication has 33 references indexed in Scilit:
- Modeling the Cost Savings of Continuous Pulse Oximetry and Capnography Monitoring of United States General Care Floor Patients Receiving Opioids Based on the PRODIGY TrialAdvances in Therapy, 2021
- Postoperative respiratory state assessment using the Integrated Pulmonary Index (IPI) and resultant nurse interventions in the post-anesthesia care unit: a randomized controlled trialJournal of Clinical Monitoring and Computing, 2020
- Acoustic respiration rate and pulse oximetry-derived respiration rate: a clinical comparison studyJournal of Clinical Monitoring and Computing, 2018
- Twenty‐four‐hour continuous and remote monitoring of respiratory rate using a medical radar system for the early detection of pneumonia in symptomatic elderly bedridden hospitalized patientsClinical Case Reports, 2018
- Continuous remote monitoring of COPD patients—justification and explanation of the requirements and a survey of the available technologiesMedical & Biological Engineering & Computing, 2018
- Importance of respiratory rate for the prediction of clinical deterioration after emergency department discharge: a single‐center, case–control studyAcute Medicine & Surgery, 2016
- Sleep and BreathingClinics in Chest Medicine, 2014
- Continuous Monitoring in an Inpatient Medical-Surgical Unit: A Controlled Clinical TrialThe American Journal of Medicine, 2014
- Respiration rate monitoring methods: A reviewPediatric Pulmonology, 2011
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986