Open-Source Technology for Real-Time Continuous Glucose Monitoring in the Neonatal Intensive Care Unit: Case Study in a Neonate With Transient Congenital Hyperinsulinism (Preprint)
Authors:
- Katarina BrauneKBKatarina Braune
- Mandy WäldchenMWMandy Wäldchen
- Klemens RaileKRKlemens Raile
- Sigrid HahnSHSigrid Hahn
- Tebbe UbbenTUTebbe Ubben
- Susanne RömerSRSusanne Römer
- Daniela HoeberDHDaniela Hoeber
- Nora Johanna ReibelNRNora Johanna Reibel
- Michael LaunspachMLMichael Launspach
- Oliver BlankensteinOBOliver Blankenstein
- Christoph BührerCBChristoph Bührer
Preprint
- 12 July 2020
- preprint
- Published by JMIR Publications Inc. in JMIR Preprints
Abstract
BACKGROUND Use of real-time continuous glucose monitoring (rtCGM) systems has been shown to be a low-pain, safe, and effective method of preventing hypoglycemia and hyperglycemia in people with diabetes of various age groups. Evidence on rtCGM use in infants and in patients with conditions other than diabetes remains limited. OBJECTIVE This case study describes the off-label use of rtCGM and the use of an open-source app for glucose monitoring in a newborn with prolonged hypoglycemia secondary to transient congenital hyperinsulinism during the perinatal period. METHODS The Dexcom G6 rtCGM system (Dexcom, Inc) was introduced at 39 hours of age. Capillary blood glucose checks were performed regularly. In order to benefit from customizable alert settings and detect hypoglycemic episodes, the open-source rtCGM app xDrip+ was introduced at 9 days of age. RESULTS Time in range (45-180 mg/dL) for interstitial glucose remained consistently above 90%, whereas time in hypoglycemia (<45 mg/dL) decreased. Mean glucose was maintained above 70 mg/dL at 72 hours of life and thereafter. Daily sensor glucose profiles showed cyclic fluctuations that were less pronounced over time. CONCLUSIONS While off-label use of medication is both common practice and a necessity in newborn infants, there are few examples of off-label uses of medical devices, rtCGM being a notable exception. Real-time information allowed us to better understand glycemic patterns and to improve the quality of glycemic control accordingly. Severe hypoglycemia was prevented, and measurement of serum levels of insulin and further lab diagnostics were performed much faster, while the patient’s individual burden caused by invasive procedures was reduced. Greater customizability of threshold and alert settings would be beneficial for user groups with glycemic instability other than people with diabetes, and for hospitalized newborn infants in particular. Further research in the field of personal and off-label rtCGM use, efficacy studies evaluating the accuracy of low glucose readings, and studies on the differences between algorithms in translating raw sensor data, as well as customization of commercially available rtCGM systems, is needed.Keywords
All Related Versions
- Published version: Journal of Medical Internet Research, 22 (12), e21770.
This publication has 27 references indexed in Scilit:
- Feasibility of automated insulin delivery guided by continuous glucose monitoring in preterm infantsArchives of Disease in Childhood - Fetal and Neonatal Edition, 2019
- Procedural Pain during Insertion of a Continuous Glucose Monitoring Device in Preterm InfantsThe Journal of Pediatrics, 2018
- Congenital Hyperinsulinism: Diagnosis and Treatment UpdateJournal of Clinical Research in Pediatric Endocrinology, 2018
- Congenital HyperinsulinismPediatric Annals, 2017
- Continuous glucose monitoring in neonates: a reviewMaternal Health, Neonatology and Perinatology, 2017
- Empfehlungen zu Diagnostik und Therapie des kongenitalen Hyperinsulinismus (KHI)2015
- Clinically stable very low birthweight infants are at risk for recurrent tissue glucose fluctuations even after fully established enteral nutritionArchives of Disease in Childhood - Fetal and Neonatal Edition, 2014
- Congenital hyperinsulinismEarly Human Development, 2010
- Management of hyperglycaemia in the preterm infantArchives of Disease in Childhood - Fetal and Neonatal Edition, 2010
- The use of a continuous glucose monitoring system in hypoglycemic disorders.Journal of Pediatric Endocrinology and Metabolism, 2004