Computational Model of Device-Induced Thrombosis and Thromboembolism
- 1 June 2005
- journal article
- research article
- Published by Springer Nature in Annals of Biomedical Engineering
- Vol. 33 (6) , 780-797
Abstract
A numerical model of thrombosis/thromboembolism (T/TE) is presented that predicts the progression of thrombus growth and thromboembolization in low-shear devices (hemodialyzers, oxygenators, etc.). Coupled convection–diffusion-reaction equations were solved to predict velocities, platelet agonist (ADP, thromboxane A2, and thrombin) concentrations, agonist-induced and shear-induced platelet activation, and platelet transport and adhesion to biomaterial surfaces and adherent platelets (hence, thrombus growth). Single-platelet and thrombus embolization were predicted from shear forces and surface adhesion strengths. Values for the platelet-biomaterial reaction constant and the platelet adhesion strength were measured in specific experiments, but all other parameter values were obtained from published sources. The model generated solutions for sequential time steps, while adjusting velocity patterns to accommodate growing surface thrombi. Heparinized human blood was perfused (0.75 ml/min) through 580 μm-ID polyethylene flow cells with flow contractions (280 μm-ID). Thrombus initiation, growth, and embolization were observed with videomicroscopy, while embolization was confirmed by light scattering, and platelet adhesion was determined by scanning electron microscopy. Numerical predictions and experimental observations were similar in indicating: 1) the same three thrombotic locations in the flow cell and the relative order of thrombus development in those locations, 2) equal thrombus growth rates on polyethylene and silicon rubber (in spite of differing overall T/TE), and 3) similar effects of flow rate (1.5 ml/min versus 0.75 ml/min) on platelet adhesion and thrombosis patterns.Keywords
This publication has 37 references indexed in Scilit:
- Computational Flow Modeling in Hollow‐Fiber DialyzersArtificial Organs, 2002
- Impact of oxygenator design on hemolysis, shear stress, and white blood cell and platelet countsJournal of Cardiothoracic and Vascular Anesthesia, 1996
- Dynamics of platelet glycoprotein IIb-IIIa receptor expression and fibrinogen binding. I. Quantal activation of platelet subpopulations varies with adenosine diphosphate concentrationBiophysical Journal, 1994
- Dynamic measurements of the platelet membrane glycoprotein IIb-IIIa receptor for fibrinogen by flow cytometry. I. Methodology, theory and results for two distinct activatorsBiophysical Journal, 1991
- Platelet adherence and detachment: A flow study with a series of hydroxyethyl methacrylate-ethyl methacrylate copolymers using video microscopyJournal of Biomedical Materials Research, 1991
- Mathematical analysis of mural thrombogenesis. Concentration profiles of platelet-activating agents and effects of viscous shear flowBiophysical Journal, 1989
- Embolization: Critical thrombus height, shear rates, and pulsatility. Patency of blood vesselsJournal of Biomedical Materials Research, 1989
- Rheological Aspects of Thrombosis and Haemostasis: Basic Principles and ApplicationsThrombosis and Haemostasis, 1986
- Fluid shear as a possible mechanism for platelet diffusivity in flowing bloodJournal of Biomechanics, 1986
- Biodynamics1984