Behavior of leadless AV synchronous pacing during atrial arrhythmias and stability of the atrial signals over time—Results of the MARVEL Evolve subanalysis
- 27 January 2019
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 42 (3) , 381-387
Abstract
Introduction: The MARVEL study demonstrated at a single time point that accelerometer (ACC)‐based atrial sensing improves atrioventricular (AV) synchrony (AVS) in patients with AV block and a Micra pacemaker (Medtronic, Minneapolis, MN, USA). The purpose of the MARVEL Evolve substudy was to assess the performance over time.Methods: This prospective single‐center study compared AVS and ACC signals at two visits ≥6 months apart. Custom software was temporarily downloaded into the Micra at each visit and AVS was measured during 30 min at rest.Results: Nine patients from the MARVEL study were enrolled. The mean (±standard deviation) age was 82.3 ± 6.0 years old, 67% were male, and a Micra was implanted for 6.0 ± 6.4 months. High‐degree AV block was present in four patients, whereas five with predominantly intrinsic conduction required intermittent pacing for bradycardia. The mean interval between visits was 7.1 ± 0.6 months. Seven patients had normal sinus node function at both visits and were included in a paired analysis. Both ACC signal amplitude (visit 2‐visit 1 = 1.4 mG; 95% confidence interval [CI] [−25.8 to 28.4 mG]; P = 0.933) and AVS (visit 1: 90.8%, 95% CI [72.4, 97.4] and visit 2: 91.4%, 95% CI [63.8, 98.5]; P = 0.740) remained stable. Three patients had spontaneous atrial tachycardia. During atrial fibrillation, no atrial contraction was detected or tracked. During atrial flutter, intermittent tracking resulted in a ventricular rate of 60 ± 8 beats per minute (bpm); there was no ventricular pacing >100 bpm.Conclusion: ACC signals amplitude and performance of AVS pacing were stable over time. During atrial arrhythmias, the AV synchronous pacing mode behaved safely.Keywords
Funding Information
- Medtronic
This publication has 12 references indexed in Scilit:
- Response to atrial arrhythmias in an atrioventricular synchronous ventricular leadless pacemaker: A case report in a paroxysmal atrial fibrillation patientHeartRhythm Case Reports, 2018
- Accelerometer-based atrioventricular synchronous pacing with a ventricular leadless pacemaker: Results from the Micra atrioventricular feasibility studiesHeart Rhythm, 2018
- Detection of new atrial fibrillation in patients with cardiac implanted electronic devices and factors associated with transition to higher device-detected atrial fibrillation burdenHeart Rhythm, 2018
- A leadless pacemaker in the real-world setting: The Micra Transcatheter Pacing System Post-Approval RegistryHeart Rhythm, 2017
- Long-term performance of a transcatheter pacing system: 12-Month results from the Micra Transcatheter Pacing StudyHeart Rhythm, 2017
- A Leadless Intracardiac Transcatheter Pacing SystemNew England Journal of Medicine, 2016
- 2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: The Task Force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA)EP Europace, 2013
- Trends in Permanent Pacemaker Implantation in the United States From 1993 to 2009JACC, 2012
- Interatrial Mechanical Dyssynchrony Worsened Atrial Mechanical Function in Sinus Node Disease With or Without Paroxysmal Atrial FibrillationJournal of Cardiovascular Electrophysiology, 2009
- Quality of Life and Clinical Outcomes in Elderly Patients Treated with Ventricular Pacing as Compared with Dual-Chamber PacingNew England Journal of Medicine, 1998