Charge Storage: Stability measures in implantable electrodes
- 1 September 2009
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE) in 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
- Vol. 2009 (1094687X) , 658-661
Abstract
Here we report on long-term (300 to 600 hours) stability measures for implantable stimulating electrodes. We have considered several measures of stability as they refer to reliability of charge carrying capacity in implantable electrodes. We have designed and manufactured coatings for large area (1 to 2 mm2) stainless steel substrates. Materials tested were electrodeposited iridium oxide films, multi-walled carbon nanotube mesh, and PEDOT:PSS. Traditional characterization techniques such as cyclic voltammetry and electrochemical impedance spectroscopy cover a small fraction of the characterization framework needed for ensuring the safety and performance of electrodes designed for long-term implants. The stability measures suggested here rely on continuous low frequency cycling and evaluation of cathodic charge storage capacity during cycling. We experimentally show, in this paper, that the stability may be measured and is relevant for long-term applications of such coatings.Keywords
This publication has 7 references indexed in Scilit:
- Electrochemical characterization of multi-walled carbon nanotube coated electrodes for biological applicationsCarbon, 2009
- The influence of electrolyte composition on thein vitrocharge-injection limits of activated iridium oxide (AIROF) stimulation electrodesJournal of Neural Engineering, 2007
- Electrical stimulation of excitable tissue: design of efficacious and safe protocolsJournal of Neuroscience Methods, 2005
- Microporous conducting polymers on neural microelectrode arraysSensors and Actuators A: Physical, 2004
- Five-Year Follow-up of Bilateral Stimulation of the Subthalamic Nucleus in Advanced Parkinson's DiseaseNew England Journal of Medicine, 2003
- Electrodeposited iridium oxide for neural stimulation and recording electrodesIEEE Transactions on Neural Systems and Rehabilitation Engineering, 2001
- Nonlinear transient response of electrode—electrolyte interfacesMedical & Biological Engineering & Computing, 2000